Skip to the content

FDA Medical Device Recalls (openFDA)

Edwards Lifesciences, LLC: Edwards Lifesciences Duraflo coated Femoral Cannula. Product Usage: The cannulae are intended for use in situations which require rapid femoral venous and arterial access for short-term (< 6 hours) cardiopulmonary bypass. Vessel access (venous or arterial) is left to the discretion of the physician. Extracorporeal circuit components with a Duraflo coating are intended for use in cardiopulmonary surgery when a heparin-coated blood path is desired. The femoral arterial cannulae are wire-reinforced, thin-wall cannulae.

Recall number
Z-1111-2013
Recalling firm
Edwards Lifesciences, LLC
Product
Edwards Lifesciences Duraflo coated Femoral Cannula. Product Usage: The cannulae are intended for use in situations which require rapid femoral venous and arterial access for short-term (< 6 hours) cardiopulmonary bypass. Vessel access (venous or arterial) is left to the discretion of the physician. Extracorporeal circuit components with a Duraflo coating are intended for use in cardiopulmonary surgery when a heparin-coated blood path is desired. The femoral arterial cannulae are wire-reinforced, thin-wall cannulae.
Status
Terminated
Initiated
2013-03-14
Posted
2013-04-14
Terminated
2016-03-08
Reason
Two unsealed pouches were found at an Edwards' distribution location in Japan. Incorrectly sealed pouches may result in a sterility barrier breach.
Root cause
Packaging process control
Action
Edwards Lifesciences sent an Urgent Field Safety Notice letter dated March 28, 2013 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to discontinue use of any affected product remaining in inventory and to return any unused devices to Edwards. A completed Acknowledgement Form should be faxed to Edwards at 800-422-9329 within three days of receiving the notice. Questions regarding the recall should be directed to Edwards Customer Services at 800-424-3278, 6:00AM - 4:30PM PST.
Quantity
379 units
Distribution
USA Nationwide Distribution including the states of: FL, MA, MI, NY, PA, TX.
Lot, serial or model codes
Model DIIFEMII018A, Lot 59337174
510(k) numbers
["K123298"]
PMA numbers
not on file
Product code
DWF
Firm FEI
1713910
Firm city
Draper
Firm state
UT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register