Skip to the content

FDA Medical Device Recalls (openFDA)

Philips North America Llc: IntelliVue Patient Monitor MX500. Model Number: 866064

Recall number
Z-1109-2025
Recalling firm
Philips North America Llc
Product
IntelliVue Patient Monitor MX500. Model Number: 866064
Status
Open, Classified
Initiated
2025-01-15
Posted
2025-02-05
Terminated
not on file
Reason
Potential failure to alarm for "No Device Data" INOP. The device to alarm after a Hamilton ventilator is disconnected from the EC10 Module or IntelliBridge I/O board.
Root cause
Software design
Action
Philips began notifying consignees on about 01/15/2025 via certified letters titled URGENT Field Safety Notice and URGENT Medical Device Correction letters. Consignees were instructed to ensure that there is no interruption of the cable connection between the Hamilton ventilator and the EC10 Module or IntelliBridge I/O board before use, provide the notification to all affected personnel and customers if the devices have been further distributed, and to complete and return the response form provided. A Philips representative will contact you to schedule a visit from a Philips Field Service Engineer who will upgrade your ED245 Driver from revision A.0 to revision A.1. Distributors were instructed to return the Response Form attached your contact information, send the attached URGENT Field Safety Notice to each customer to whom you have distributed any affected device as soon as possible and no later than 5 days, together with the Reply Card, perform a good faith effort to get the Response Form by following up with the customer with a minimum of three attempts, and if possible, using multiple contact methods, inform Philips about the responses received, and implement the Technical solution established by Philips as soon as available within the timeframe communicated by Philips and confirm to Philips the implementation in all affected products. In case a product is not corrected, inform Philips on the reason why the technical solution cannot be implemented.
Quantity
45 units
Distribution
US distribution to Florida, North Carolina, and Oregan. International distribution to Canada, Estonia, France, Germany, Japan, Netherlands, Spain, Switzerland, United Kingdom.
Lot, serial or model codes
Model Number: 866064. UDI-DI: (01)00884838038776(21); Serial Numbers: DE7580A1R5, DE7580A1R4, DE7580A1R1, DE7580A1R0, DE7580A1PM, DE7580A1PL, DE7580A1PK, DE7580A1PJ, DE7580A1P5, DE351C8913, DE351C8912, DE351C8908, DE351C8911, DE351C8910, DE351C8909, DE32300881, DE32300882, DE32300883, DE32300937, DE32300936, DE32301099, DE35114607, DE35114606, DE671T0786, DE671K1669, DE671V9213, DE35133301, DE671N0625, DE671E0967, DE671E0964, DE671E0965, DE671E0968, DE671E0978, DE671E1009, DE671E1008, DE671E1007, DE671E1006, DE671E1005, DE671E1004, DE671E0977, DE671E1003, DE671E0979, DE671E1015, DE671E0966, DE671E1059.
510(k) numbers
["K082583"]
PMA numbers
not on file
Product code
MHX
Firm FEI
3016618143
Firm city
Cambridge
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register