Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: SIEMENS AXIOM Artis - AUTOMAP FUNCTION

Recall number
Z-1108-04
Recalling firm
Siemens Medical Solutions USA, Inc
Product
SIEMENS AXIOM Artis - AUTOMAP FUNCTION
Status
Terminated
Initiated
2004-06-02
Posted
2004-07-20
Terminated
2004-10-21
Reason
Error appears in the AUTOMAP function, Module 2, when PERISTEEPING or PERIVISION is activated simultaneously that causes system to crash.
Root cause
Other
Action
The recalling firm sent recall letter to their customers informing them of the software problem. The customers were informed that a Siemens Service Representative will visit each location to deactivate Module 2 of the AUTOMAP function per Update AX023/04S.
Quantity
127 units
Distribution
The recalled product has been distributed throughout the United States, such as: AK, AL, AZ, CA, DC, FL, GA, IA, IL, IN, KY, LA, MI, MN, MO, MS, NC, NJ, NY, OH, OK, PA, SC, TN, TX, VA, WA, WI,
Lot, serial or model codes
Model Numbers:  AXIOM Artis FC 5904433 AXIOM Artis BC 5904649 AXIOM Artis MP 5904466 AXIOM Artis FA 5904441 AXIOM Artis BA 5904656 AXIOM Artis TA 7007755  AXIOM Artis FC: 10050; 10052; 10055; 10064; 10072; 10086; 10094; 10124; 10148; 10151; 10153; 10176; 10177; 10182; 10198; 10217; 10228; 10237; 10240; 10243; 10249; 10251; 10253; 10266; 10284; 10298; 10307; 10326; 10335; 10338; 10348; 10364; 10365;  10367; 10369  AXIOM Artis BC: 14007  AXIOM Artis MP: 20022; 20024; 20032; 20036; 20045; 20048; 20070; 20078; 20082; 20090; 20094; 20107; 20110; 20111; 20155; 20169; 20177; 20190; 20199; 20203; 20205
510(k) numbers
["K010721"]
PMA numbers
not on file
Product code
IZI
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register