Skip to the content

FDA Medical Device Recalls (openFDA)

Randox Laboratories Ltd.: Liquid Immunoassay Premium Controls, Model Number LIA3105 Product Usage: This product is intended for in vitro diagnostic use, as assayed quality control material to monitor the accuracy and reproducibility of analytes listed in the package insert

Recall number
Z-1107-2019
Recalling firm
Randox Laboratories Ltd.
Product
Liquid Immunoassay Premium Controls, Model Number LIA3105 Product Usage: This product is intended for in vitro diagnostic use, as assayed quality control material to monitor the accuracy and reproducibility of analytes listed in the package insert
Status
Terminated
Initiated
2019-02-07
Posted
not on file
Terminated
2020-06-08
Reason
The target value for uptake is incorrect on the value sheet for an analyzer. The target value for this analyzer is listed as 43.8%U and should be 39.4%U.
Root cause
Employee error
Action
Urgent Medical Device Correction letters dated 2/7/19 were sent to customers. The letter identified the affected product, problem and actions to be taken. For questions contact Randox Technical Services.
Quantity
4
Distribution
The products were distributed to the following US states: WV and Puerto Rico.
Lot, serial or model codes
Lot codes: 1723EC
510(k) numbers
["K112337"]
PMA numbers
not on file
Product code
JJY
Firm FEI
1000361607
Firm city
Crumlin
Firm state
North
Firm country
Ireland

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register