FDA Medical Device Recalls (openFDA)
Zimmer Manufacturing B.V.: Product 9 consists of all product under HRS, and same usage: Item no: 47494502601 4.5 BROAD SCP PLT, 26 HOL Product Usage: Temporary internal fixation devices designed to stabilize fractures during the normal healing process.
- Recall number
- Z-1107-2016
- Recalling firm
- Zimmer Manufacturing B.V.
- Product
- Product 9 consists of all product under HRS, and same usage: Item no: 47494502601 4.5 BROAD SCP PLT, 26 HOL Product Usage: Temporary internal fixation devices designed to stabilize fractures during the normal healing process.
- Status
- Terminated
- Initiated
- 2016-01-11
- Posted
- 2016-03-12
- Terminated
- 2017-04-13
- Reason
- As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.
- Root cause
- Packaging
- Action
- Zimmer Biomet sent an Urgent Medical Device Recall Letter, dated January 11, 2016, to Distributors, Sales Representatives, Operation Managers, and Risk Managers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to 1.Review the notification and ensure that relevant personnel are aware of the contents. 2. Assist your Zimmer Biomet sales representative with the quarantine of any affected product. 3. Your Zimmer Biomet sales representative will remove the recalled product from your facility. 4. Complete and return the attached Certificate of Acknowledgment form to corporatequality.postmarket@zimmerbiomet.com. 5. If after reviewing this notification you have further questions or concerns please call Customer Service at 1-800-348-2759, or contact your Zimmer Biomet Sales Representative.
- Quantity
- 0
- Distribution
- US Nationwide Distribution
- Lot, serial or model codes
- lot no: 61889853
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- HRS
- Firm FEI
- 1000132037
- Firm city
- Mercedita
- Firm state
- not on file
- Firm country
- PR
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28