Roche Molecular Systems, Inc.: Amplilink Software Versions 1.1, 1.3, 1.4, 2.41 COBAS AmpliPrep/COBAS Amplicor HCV Test, version 2.0

Recall number
Z-1107-2007
Recalling firm
Roche Molecular Systems, Inc.
Product
Amplilink Software Versions 1.1, 1.3, 1.4, 2.41 COBAS AmpliPrep/COBAS Amplicor HCV Test, version 2.0
Status
Terminated
Initiated
2007-05-21
Posted
2007-08-02
Terminated
2010-02-18
Reason
For select COBAS AMPLICOR Tests run on the COBAS AMPLICOR Analyzer in conjunction with AMPLILINK software, a discrepancy has been identified between the onboard working reagent stability information reported by AMPLILINK Software (versions 1.1, 1.3, 1.4 and 2.41) and the stability information provided in the test kit package inserts/method manual.
Root cause
Other
Action
Urgent Medical Device Correction letters were mailed on 5/21/07 by first class mail. Letters describe the discrepancy between software reagent expiration dates and package insert expiration dates.
Quantity
5,806 total
Distribution
Nationwide to hospitals and laboratories.
Lot, serial or model codes
Catalog number 03576710190, working substrate SB3-SB
510(k) numbers
not on file
PMA numbers
["P000012S003"]
Product code
MZP
Firm FEI
2243471
Firm city
Somerville
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register