Ev3, Inc: X-Sizer Thrombectomy Catheter System (ev3) Model number XD-CS1150-45 (for US distribution). Model number XR-CS1150-45 (for foreign distribution). 1.5mm. Sterilization with Ethylene Oxide Gas. ev3 4600 Nathan Lane North, Plymouth, MN 55442-2920 USA.

Recall number
Z-1107-06
Recalling firm
Ev3, Inc
Product
X-Sizer Thrombectomy Catheter System (ev3) Model number XD-CS1150-45 (for US distribution). Model number XR-CS1150-45 (for foreign distribution). 1.5mm. Sterilization with Ethylene Oxide Gas. ev3 4600 Nathan Lane North, Plymouth, MN 55442-2920 USA.
Status
Terminated
Initiated
2006-02-14
Posted
2006-06-14
Terminated
2006-12-26
Reason
Loss of Operating Vacuum during use-It has been determined that a small percentage of X-Sizer Thrombectomy Catheter Systems from specific lots may lose operating vacuum prematurely due to a system leak within the Control Module. This failure can lead to the inaility to complete the procedure of thrombus removal and the necessity to remove the device prior to achieving a satisfactory result.
Root cause
Other
Action
A medical Device Recall Letter dated 02/13/06 was sent to hospitals along the affected device lot numbers the hospital should have received on a Device Reconciliation Form. ev3 is asking for affected devices to be returned. Returned devices will be replaced at no cost.
Quantity
Model number XD-CS1150-45 (US) 534 devices. Model number XR-CS1150-45 (OUS) 504 devices.
Distribution
World wide-AL, AR, AZ, CA, CO, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MO, MN, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV. OUS to: Argentina, Australia, Austria, Brazil, Canada, Chile, Columbia, Czech Republic, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Italy, Jordan, Mexico, Netherlands, Norway, Pakistan, Portugal, Saudi Arabi, Singapore, South Africa, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom
Lot, serial or model codes
Model number XD-CS1150-45 (for US distribution), lot numbers: 782344, 782345, 782366, 782379, 782386, 782389,  782403, 782404, 782406, 782407, 782412, 782421, 782428,  782429, 782430, 782436, 782438, 782439, 782441, 782445,  782526, 986604, 986604, 986604, 986605, 986609, 986620,  986626, 986640, 986641, 986642, 986645, 986656, 986659,  1174394, 1174395, 1212529, 1229730, 1231296, 1234273,  1240986, 1241067, 1271238, 1272506, 1280664, 1280952,  1281088, 1282693, 1289590, 1289978, 1293902, 1295372,  1300605, 1302005, 1303572, 1308536, 1310238, 1362165.   Model number XR-CS1150-45 (for OUS distribution), lot numbers: 782773, 782777, 782787, 782788, 782790, 782791,  782795, 782796, 782797, 782798, 782800, 782801, 782804,  782808, 782811, 782812, 782816, 782818, 782822, 782848,  987054, 987055, 987056, 987058, 987059, 987062, 987063,  987064, 987065, 987066, 987067, 987070, 987070, 987070,  987071, 987073, 987076, 987078, 1057871, 1077841, 1077842, 1077843, 1077844, 1077846, 1077847, 1077848,  1077854, 1077856, 1077859, 1077862, 1095087, 1095089,  1296437, 1296552, 1313297, 1313502, 1313543, 1313562,  1313611, 1313645, 1314989, 1314991, 1314994, 1315002, 1315042, 1315058, 1315113, 1315140, 1315183, 1315211,  1315245, 1315292, 1315294, 1315313, 1315328, 1315365,  1315378, 1315383, 1315623, 1315624, 1315625, 1315626,  1315632, 1315633, 1315636, 1315639, 1315640, 1315644,  1315646, 1317128, 1317133, 1317134, 1317135, 1317144,  1317301, 1326969, 1326969, 1326969, 1326970, 1326972,  1326973, 1326974, 1326977, 1326982, 1326983, 1326998,  1326999, 1327006, 1327008, 1327011, 1327546, 1342794,  1342852, 1342874, 1342882, 1342888, 1342957, 1342980,  1343007, 1343019, 1343036, 1343045, 1343094, 1343108,  1343125, 1343136, 1343266, 1343273, 1343292, 1343314,  1363701, 1363712, 1363725, 1363742, 1363791, 1363827,  1363850, 1363974, 1363976, 1364006, 1364034, 1364049.
510(k) numbers
["K021096"]
PMA numbers
not on file
Product code
MCW
Firm FEI
2183870
Firm city
Plymouth
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register