Skip to the content

FDA Medical Device Recalls (openFDA)

Medtronic Physio Control Corp: LIFEPAK 12 defibrillator/monitor Part Numbers: VLP12-02-000000 through VLP12-02-005003, AVLP12-02-000000 through AVLP12-02-005003, DVLP12-02-000000 through DVLP12-02-005003, EVLP12-02-000000 through EVLP12-02-005003, FVLP12-02-000000 through FVLP12-02-005003, LVLP12-02-000000 through LVLP12-02-005003, PVLP12-02-000000 through PVLP12-02-005003, UVLP12-02-000000 through UVLP12-02-005003

Recall number
Z-1107-04
Recalling firm
Medtronic Physio Control Corp
Product
LIFEPAK 12 defibrillator/monitor Part Numbers: VLP12-02-000000 through VLP12-02-005003, AVLP12-02-000000 through AVLP12-02-005003, DVLP12-02-000000 through DVLP12-02-005003, EVLP12-02-000000 through EVLP12-02-005003, FVLP12-02-000000 through FVLP12-02-005003, LVLP12-02-000000 through LVLP12-02-005003, PVLP12-02-000000 through PVLP12-02-005003, UVLP12-02-000000 through UVLP12-02-005003
Status
Terminated
Initiated
2004-05-05
Posted
2004-07-20
Terminated
2006-05-06
Reason
Therapy cable doesn't meet firm's standards/may not maintain connections sufficient to support ECG monitoring, external pacing or defibrillation; ECG connector with higher than normal connection resistance resulting in intermittent ECG signal; therapy connector has potential for impact damage.
Root cause
Other
Action
The firm reviewed records and/or contacted customers by telephone to verify affected devices used in the out of hospital mobile environment. All affected customers were notified by a letter dated May 2004. The firm''s service representative will contact customers to inspect and correct the devices.
Quantity
20,676 domestic, 19,899 international
Distribution
The firm distributes to distributors, hospitals, EMT-Ds and paramedics nationwide. The firm also distibutes the product internationally.
Lot, serial or model codes
Non-continuous Serial Numbers between 1178083 and 305080962.
510(k) numbers
["K012274"]
PMA numbers
not on file
Product code
MKJ
Firm FEI
3015876
Firm city
Redmond
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register