Skip to the content

FDA Medical Device Recalls (openFDA)

Arrow International Inc: Arrow AGB+ Multi-Lumen CVC Kit 7Fr 3 Lumen- Intended for venous access to the central circulation by the way of the femoral, jugular, or subclavian veins. Product Code: CDC-45703-1A

Recall number
Z-1106-2021
Recalling firm
Arrow International Inc
Product
Arrow AGB+ Multi-Lumen CVC Kit 7Fr 3 Lumen- Intended for venous access to the central circulation by the way of the femoral, jugular, or subclavian veins. Product Code: CDC-45703-1A
Status
Terminated
Initiated
2021-01-06
Posted
not on file
Terminated
2022-07-12
Reason
The lidstock / information card on the kits correctly states the product code as CDC-45703-1A. The banner card on the side of the kit may incorrectly state the product code as CDC-15703-1A and not identify the product as a coated ARROWg+ard Blue PLUS (AGB+) antimicrobial catheter which may cause a minor delay during use and placement of a coated catheter in patients with known sensitivity to chlorohexidine, silver sulfadiazine or sulfa could result in an allergic reaction
Root cause
Under Investigation by firm
Action
Arrow International issued Urgent Medical Device Correction notification letter on 6th Jan 2021 via FedEx 2-day mail.Letter states reason for recall, health risk and action to take: 1. If you have affected stock in inventory, immediately discontinue use and quarantine any products with the product codes and lot numbers listed above. 2. Inspect affected products within your control to confirm the product code on the lidstock matches the product code on the banner card of the same kit. Products confirmed to have an incorrect banner card should be returned to Arrow International. Products which properly match should not be returned. 3. If you have product with an incorrect banner card, please complete the enclosed Acknowledgement Form and fax it to 1-855-419-8507 Attn: Customer Service or email to recalls@teleflex.com. This will allow us to document the amount of product you have on hand for return. A customer service representative will contact you with a Return Goods Authorization (RGA) number and provide instructions for the return of products. 4. If you have no affected stock, please complete the enclosed Acknowledgement Form and fax it to 1-855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. This will allow us to document your receipt of this letter. If you have any other questions, feel free to contact your local sales representative or Customer Service at 1-866-396-2111.
Quantity
1,705 eaches
Distribution
US Nationwide
Lot, serial or model codes
Lot Code: 13F20E0173
510(k) numbers
["K993691"]
PMA numbers
not on file
Product code
FOZ
Firm FEI
3015859709
Firm city
Reading
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register