FDA Medical Device Recalls (openFDA)
Roche Molecular Systems, Inc.: Amplilink Software versions 1.1, 1.3, 1.4, 2.41. COBAS Amplicor CTA & NGA Tests
- Recall number
- Z-1106-2007
- Recalling firm
- Roche Molecular Systems, Inc.
- Product
- Amplilink Software versions 1.1, 1.3, 1.4, 2.41. COBAS Amplicor CTA & NGA Tests
- Status
- Terminated
- Initiated
- 2007-05-21
- Posted
- 2007-08-02
- Terminated
- 2010-02-18
- Reason
- For select COBAS AMPLICOR Tests run on the COBAS AMPLICOR Analyzer in conjunction with AMPLILINK software, a discrepancy has been identified between the onboard working reagent stability information reported by AMPLILINK Software (versions 1.1, 1.3, 1.4 and 2.41) and the stability information provided in the test kit package inserts/method manual.
- Root cause
- Other
- Action
- Urgent Medical Device Correction letters were mailed on 5/21/07 by first class mail. Letters describe the discrepancy between software reagent expiration dates and package insert expiration dates.
- Quantity
- 5,806 total
- Distribution
- Nationwide to hospitals and laboratories.
- Lot, serial or model codes
- Catalog number 20757497122, working probe suspension /IC PS1-IC4; Catalog Number 20757535122, Working substrate SB3-SB
- 510(k) numbers
- ["K964506"]
- PMA numbers
- not on file
- Product code
- JJF
- Firm FEI
- 2243471
- Firm city
- Somerville
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28