Skip to the content

FDA Medical Device Recalls (openFDA)

Roche Molecular Systems, Inc.: cobas SARS-CoV-2 & Influenza A/B Qualitative nucleic acid test for use on the cobas 5800/6800/8800 Systems; Material Number 09446125190

Recall number
Z-1105-2023
Recalling firm
Roche Molecular Systems, Inc.
Product
cobas SARS-CoV-2 & Influenza A/B Qualitative nucleic acid test for use on the cobas 5800/6800/8800 Systems; Material Number 09446125190
Status
Open, Classified
Initiated
2022-12-21
Posted
2023-02-07
Terminated
not on file
Reason
The firm received customer complaints regarding false negative Influenza A (Flu A) results and late Flu A Target Ct values. A false negative influenza A result may lead to additional testing, psychological distress, and delays to targeted therapy for influenza as well as true diagnosis.
Root cause
Device Design
Action
On December 22, 2022, Roche posted the Urgent Medical Device Correction (UMDC) to its website (dialog.roche.com). On December 27, 2022, the firm also sent UMDC letters to affected customers. Consignees were instructed to monitor for negative influenza A results that are inconsistent with clinical presentation and/or other clinical and epidemiological information. Authorized or licensed Influenza NAATs are available for confirmation if clinically indicated for at-risk patients. Additionally, consignees are reminded to review the tests Instructions for Use, specifically the Intended Use Statements and Procedural Limitations sections, which provide guidance on how negative results should be utilized and mutations within the target regions of the tests can impact detection, respectively. Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions about this recall
Quantity
9,948 OUS only
Distribution
Domestic distribution nationwide. Foreign distribution worldwide.
Lot, serial or model codes
UDI-DI 00875197006827 All in-date lots
510(k) numbers
not on file
PMA numbers
not on file
Product code
QLT
Firm FEI
2243471
Firm city
S Branchburg
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register