Skip to the content

FDA Medical Device Recalls (openFDA)

Intuitive Surgical, Inc.: Da Vinci Xi/X EndoWrist Monopolar Curved scissors (MCS), model number 470179-19, UDI: (00) 886874112298

Recall number
Z-1105-2020
Recalling firm
Intuitive Surgical, Inc.
Product
Da Vinci Xi/X EndoWrist Monopolar Curved scissors (MCS), model number 470179-19, UDI: (00) 886874112298
Status
Terminated
Initiated
2018-12-04
Posted
not on file
Terminated
2022-08-24
Reason
This recall is being initiated because lots of EndoWrist Monopolar Curved Scissor instruments are not recognized by the integrated electrosurgical unit, which prohibits the activation of monopolar energy from the surgeon console.
Root cause
Nonconforming Material/Component
Action
On 12/4/2018 date, the firm provided consignees a "Urgent Medical Device Recall" letter. The letter informed customers about the recall. In addition to providing information about the recall, the letter asked customers to do the following: 1.Locate and return all affected instruments in your inventory via the standard Intuitive RMA process as soon as possible. 2. Inform affected personnel when the return has been completed. 3. Please log into the da Vinci Online Community Field Action resource to read or complete any requested actions related to this issue, at this link: https://www.davincisurgerycommunity.com/ 4. In the case where the da Vinci online resource cannot be used, complete the attached Acknowledgement Form and return it via fax to Intuitive Surgical as instructed on the form. 5. Please retain a copy of this letter and the acknowledgement form for your files. 6.If you need further information or support concerning this Medical Device Recall, please contact your Clinical Sales Representative or contact Intuitive Surgical Customer Service at the numbers listed below: North and South America: (800) 876-1310, Option 3 (4 AM to 5 PM PST) or mail: customerservice@intusurg.com.
Quantity
286 Monopolar curved scissors
Distribution
US National Distribution Alabama, Arizona, California, Colorado, Connecticut, Delaware, Florida, Georgia, Illinois, Indiana, Iowa, Kentucky, Louisiana, Maryland, Massachusetts, Michigan, Minnesota, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New York, North Carolina, Ohio, Oklahoma, Oregon, Pennsylvania, Tennessee, Texas, Vermont, Virginia, Washington, West Virginia, Wisconsin, Wyoming OUS:None
Lot, serial or model codes
Serial Number/Manufacturing Lot: N11181003, N11181004
510(k) numbers
["K131861"]
PMA numbers
not on file
Product code
NAY
Firm FEI
3001675293
Firm city
Sunnyvale
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register