Skip to the content

FDA Medical Device Recalls (openFDA)

Smith & Nephew, Inc.: smith&nephew, LEGION HK FEMORAL WEDGE (WEDGE AND SCREW), 5 MM, SIZE 5, REF 71422159, QTY: (1), STERILE The Revision Knee System Components are indicated for rheumatoid arthritis; post-traumatic arthritis, osteoarthritis, or degenerative arthritis in older patients whose age, weight, and activity levels are compatible with an adequate long-term result; failed osteotomies, unicompartmental replacement, or total knee replacement. The constrained knee systems are designed for use in patients in primary and revision surgery, where the posterior cruciate ligament and one or both of the collateral ligaments (i.e., medial collateral and/or lateral collateral ligament) are absent or incompetent. The Revision Knee Components are for single use only and are intended for implantation with bone cement.

Recall number
Z-1105-2018
Recalling firm
Smith & Nephew, Inc.
Product
smith&nephew, LEGION HK FEMORAL WEDGE (WEDGE AND SCREW), 5 MM, SIZE 5, REF 71422159, QTY: (1), STERILE The Revision Knee System Components are indicated for rheumatoid arthritis; post-traumatic arthritis, osteoarthritis, or degenerative arthritis in older patients whose age, weight, and activity levels are compatible with an adequate long-term result; failed osteotomies, unicompartmental replacement, or total knee replacement. The constrained knee systems are designed for use in patients in primary and revision surgery, where the posterior cruciate ligament and one or both of the collateral ligaments (i.e., medial collateral and/or lateral collateral ligament) are absent or incompetent. The Revision Knee Components are for single use only and are intended for implantation with bone cement.
Status
Terminated
Initiated
2018-02-01
Posted
not on file
Terminated
2018-10-22
Reason
The screws package with devices within lot 14AAP6352A were 5 MM instead of 10MM and are too short to be used with the Legion HK Distal Wedge Size 5.
Root cause
Process control
Action
The firm initiated their recall by email on February 1, 2018. The letter stated the following: "Required Actions: 1. Please inspect your inventory and locate any unused devices from the listed product and batch numbers on the first page of this Field Action Notification, and quarantine them immediately. a. If you are a distributor, you must notify your customers of the field action and ensure that these actions are carried out. 2. If you have no product to return, please put an X in the appropriate location below. 3. If you have product to return, please list the batch numbers and quantities of each batch that you are returning in the appropriate boxes below. 4. Complete the remainders of the form sign and send to FieldActions@smith-nephew.com or fax to 901-566- 7975. Please Note even if you have no product to return, this form must be completed, signed and returned. 5. Once the form is received by Smith & Nephew, you will be sent a Return Authorization (RA) number. If you have any questions or concerns regarding this recall please contact FieldActions@smith-nephew.com." For further questions, please call (901) 396-2121.
Quantity
28 units
Distribution
Worldwide Distribution - US Distribution to the states of: NY, IL, NJ, TX, CO, and to the countries of : Belgium, Great Britain, Australia, Germany, Netherlands, Norway.
Lot, serial or model codes
Batch Number14AAP6352A
510(k) numbers
["K081111"]
PMA numbers
not on file
Product code
KRO
Firm FEI
1020279
Firm city
Memphis
Firm state
TN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register