Skip to the content

FDA Medical Device Recalls (openFDA)

Smith & Nephew, Inc Endoscopy Division: Smith & Nephew TRIVEX System Pneumatic Footswitch, Part #7209791; distributed by Smith & Nephew, Inc. - Endoscopy, Andover, MA 01810.

Recall number
Z-1105-2007
Recalling firm
Smith & Nephew, Inc Endoscopy Division
Product
Smith & Nephew TRIVEX System Pneumatic Footswitch, Part #7209791; distributed by Smith & Nephew, Inc. - Endoscopy, Andover, MA 01810.
Status
Terminated
Initiated
2004-10-12
Posted
2007-08-02
Terminated
2007-08-02
Reason
Build-up of static charge in Pneumatic Footswitch tubing may cause control unit to reset to default settings in the operating room during surgery.
Root cause
Other
Action
Firm sent notification letters to consignees via overnight FedEx on 10/12-13/04. International customers were notified via letter on 11/10/04. Customers were asked to return footswitches for a replacement.
Quantity
274 units.
Distribution
Product sent to distributors nationwide and to the following countries: Australia, Austria, Canada, China, Germany, Italy, Japan, and Taiwan.
Lot, serial or model codes
Pnuematic Footswitch part of TRIVEX System with the following serial numbers: 105, 106, 108, 109, 111 -123, 125 - 182, 184 - 188, 192 - 197, 199 - 203, 205 - 219, 222 - 249, 252 - 254, and 7210386.
510(k) numbers
["K032387"]
PMA numbers
not on file
Product code
DWQ
Firm FEI
1643264
Firm city
Oklahoma City
Firm state
OK
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register