Skip to the content

FDA Medical Device Recalls (openFDA)

Roche Diagnostics Corp.: Disetronic D-Tronplus Insulin Pump adapters; Ref 3000803; catalog reference number 0457243001 sold in packages of ten adapters.

Recall number
Z-1105-05
Recalling firm
Roche Diagnostics Corp.
Product
Disetronic D-Tronplus Insulin Pump adapters; Ref 3000803; catalog reference number 0457243001 sold in packages of ten adapters.
Status
Terminated
Initiated
2005-07-15
Posted
2005-08-06
Terminated
2006-02-07
Reason
The adapter valve may not operate correctly, which may result in overdelivery of insulin in the first dose after the adapter has been changed.
Root cause
Other
Action
The firm began notifying family members of pediatric patients via phone on 7/15/05 and immediately began shipping a recall letter and 10 packs of replacement adapters to each pediatric user. Attending physicians were for these pediatric patients were notified the week of 7/18/05 via phone and follow-up letters were sent to these physicians. The firm is issuing a press release on 7/28/05. Only these replacement adapters are being sold by Disetronic to users of any age ordering new adapters.
Quantity
107 pediatric patients use the device.
Distribution
Nationwide.
Lot, serial or model codes
All units with the following lot numbers being used by patients 13 years of age or younger: lots 4013674 through 4022628. The last affected lot expires in April 2007.
510(k) numbers
["K043000"]
PMA numbers
not on file
Product code
LZG
Firm FEI
1823260
Firm city
Indianapolis
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register