Dexcom Inc: Brand Name: Dexcom G6 CGM App for iOS Software Number: SW11677 Software Version: Version 1.4.0 up to corrected Version 1.4.4 GUDID Registered GTIN: 00386270000804

Recall number
Z-1104-2020
Recalling firm
Dexcom Inc
Product
Brand Name: Dexcom G6 CGM App for iOS Software Number: SW11677 Software Version: Version 1.4.0 up to corrected Version 1.4.4 GUDID Registered GTIN: 00386270000804
Status
Terminated
Initiated
2019-11-19
Posted
not on file
Terminated
2023-08-22
Reason
It was reported that the user's low alarm feature on the iOS application were not properly alerting users when the user has enabled the Alert Schedule feature more than 30 days after installing the iOS application.
Root cause
Software design
Action
An updated G6 iOS CGM App was placed in the Apple App Store (SW11677 version 1.4.4) with messaging pushed to all US users on November 19, 2019. The messaging repeats to the user on a daily basis until the user updates (or is alternatively blocked from use of the app if they ignore the messaging and do not update). Messaging will be delivered directly to the end user through the following App Compatibility Messaging via the Apple App Store (see Compatibility Message and time-frame below): Duration: Day 1-3 Date Range: Nov 19-21 App Compatibility Message: Warning - Please update your Dexcom G6 iOS App from the Apple App Store. Failure to update your App may cause your alarms and alerts to not work as intended. Without alarms/alerts, you might miss a severe low or high glucose event. Contact technical support if you require assistance. Duration: Day 4-17 Date Range: Nov 22 - Dec 5 App Compatibility Message: Warning - This version of the Dexcom G6 iOS App will no longer be available for use as of December 6. Update the Dexcom G6 iOS App now from the Apple App Store to continue use of the Dexcom G6 iOS App. Contact technical support if you require assistance. Duration: Day 18 Date Range: Dec 6 App Compatibility Message: Warning - This version of the Dexcom G6 iOS App is no longer available for use. Update the Dexcom G6 iOS App now from the Apple App Store to continue use of the Dexcom G6 iOS App. Contact technical support if you require assistance. For impacted users, they will receive the Alert Settings Reset notification once the app update is complete and the app is launched or the app receives a communication from the Transmitter. The user must then reset any alarm/alert customizations made (or leave at factory default settings). For users currently not impacted, they will not receive the Alert Settings Reset notification. The updated version of the app corrects the problem for those impacted and prevents the problem from occurring for
Quantity
237888 users registered.
Distribution
US Nation-wide OUS: Austria; Belgium; Canada; Czechia; Denmark; Finland; Germany; Great Britain; Hong Kong; Ireland; Italy; Luxembourg; Netherlands; Norway; South Africa; Spain; Sweden, and Switzerland.
Lot, serial or model codes
Software Number: SW11677 Software Version: 1.4.0 up to corrected Version 1.4.4 GUDID Registered GTIN: 00386270000804
510(k) numbers
["K191450"]
PMA numbers
not on file
Product code
QBJ
Firm FEI
3004753838
Firm city
San Diego
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register