DePuy Orthopaedics, Inc.: Universal Femoral Sleeves - Product Usage: used primarily in revision surgeries with the S-ROM Noiles, LPS, SIGMA, and LCS Knee products.

Recall number
Z-1103-2021
Recalling firm
DePuy Orthopaedics, Inc.
Product
Universal Femoral Sleeves - Product Usage: used primarily in revision surgeries with the S-ROM Noiles, LPS, SIGMA, and LCS Knee products.
Status
Terminated
Initiated
2021-01-08
Posted
not on file
Terminated
2024-01-17
Reason
The taper dimensions may be out of specification due to a production issue during the manufacturing of these lots.
Root cause
Nonconforming Material/Component
Action
On 01/12/2021, DePuy Synthes issued an Urgent Medical Device Recall (removal) via letter due to taper dimensions of certain Universal Femoral Sleeves may be out of specification. The Universal Femoral Sleeves are used primarily in revision surgeries with the S-ROM Noiles, LPS, SIGMA, and LCS Knee products. DePuy is recommending that customers return the recalled devices.
Quantity
60 sleeves
Distribution
Worldwide distribution - US Nationwide distribution in the states of AK, CA, FL, IN, MI, MN, MS, NC, NY, OH, PA, TX, VA and the countries of Canada, China, Germany, Puerto Rico, Switzerland.
Lot, serial or model codes
Product No.: 129453216, 129453226, 129453236,129453246 Lot No.: J85Y90, J86T23, J86T20, J8790J, J85Y92
510(k) numbers
["K063633"]
PMA numbers
not on file
Product code
JWH
Firm FEI
1818910
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register