Skip to the content

FDA Medical Device Recalls (openFDA)

RAYSEARCH LABORATORIES AB: RayStation/RayPlan. Model/Catalog Number: 10.1. Software Version: RayStation/RayPlan 10B. Radiation Therapy Treatment Planning System

Recall number
Z-1102-2026
Recalling firm
RAYSEARCH LABORATORIES AB
Product
RayStation/RayPlan. Model/Catalog Number: 10.1. Software Version: RayStation/RayPlan 10B. Radiation Therapy Treatment Planning System
Status
Open, Classified
Initiated
2025-11-28
Posted
2026-01-16
Terminated
not on file
Reason
Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.
Root cause
Under Investigation by firm
Action
RaySearch Laboratories notified consignees via email on about 12/05/2025. Consignees were instructed to follow the outlined steps in the notification letter when adding geometry to an empty ROI with MO or type Bolus, Fixation or Support, educate planning staff and all users about the workaround, inspect units and identify all installed units with the affected software version numbers, and confirm receipt and understanding of the customer notification by responding to the notification email. The issue will be resolved in the next version of RayStation, scheduled for market release in December 2025 (subject to market clearance in some markets). If customers wish to continue using versions of RayStation affected by this notice, all users must maintain awareness of this notice. Alternatively, customers can choose to upgrade to the new version once it becomes available for clinical use. The upgrade to an unaffected software version will be scheduled in coordination with the authorized personnel, considering the specific circumstances and requirements of each individual case. The upgrade may be conducted remotely or onsite, as deemed appropriate.
Quantity
0
Distribution
Refer to the attached documents: Distribution list 167168 US - updated, Distribution list 167168 OUS
Lot, serial or model codes
UDI: 0735000201031020201216. GTIN: 07350002010310. Serial Numbers: 10.1.0.613. Software Revision: RayStation 10B. Expiration Date: 2027-02-01.
510(k) numbers
["K210645"]
PMA numbers
not on file
Product code
MUJ
Firm FEI
3007774465
Firm city
Stockholm
Firm state
not on file
Firm country
Sweden

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.