Skip to the content

FDA Medical Device Recalls (openFDA)

Wishbone Medical, Inc.: WishBone Medical Smart Correction Dual Joint Express Struts, external skeleton fixation device Type Part Number XXS 10-702-080095-0 XS 10-702-095115-0 Short 10-702-115150-0 Medium 10-702-140200 Long 10-702-190300-0

Recall number
Z-1102-2023
Recalling firm
Wishbone Medical, Inc.
Product
WishBone Medical Smart Correction Dual Joint Express Struts, external skeleton fixation device Type Part Number XXS 10-702-080095-0 XS 10-702-095115-0 Short 10-702-115150-0 Medium 10-702-140200 Long 10-702-190300-0
Status
Open, Classified
Initiated
2023-01-04
Posted
2023-02-07
Terminated
not on file
Reason
Firm identified that user may be unable to adjust the length of the Smart Correction Dual Joint Express Struts due to the collar release pin backing out. This may lead to a delay in correction.
Root cause
Process design
Action
The firm visited the affected consignee on December 30, 2021, to discuss the product issue.
Quantity
195
Distribution
US Nationwide distribution in the state of South Carolina.
Lot, serial or model codes
XXS, Part No. 10-702-080095-0 UDI-DI B5621070208009500 Lot 1048176, 1048221 XS, Part No. 10-702-095115-0 UDI-DI B5621070209511500 Lot 1048177 Short, Part No. 10-702-115150-0 UDI-DI B5621070211515000 Lot 1048178, 1048222, 1048223 Medium, Part No. 10-702-140200 UDI-DI B562107021402000 Lot 1048179 Long, Part No. 10-702-190300-0 UDI-DI B5621070219030000 Lot 1048180
510(k) numbers
["K193368"]
PMA numbers
not on file
Product code
KTT
Firm FEI
3013680140
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register