Skip to the content

FDA Medical Device Recalls (openFDA)

Baxter Healthcare Corporation: Baxter ONE-LINK Non-DEHP Y-Type Microbore Catheter Extension Set, Bonded Needle-free IV Connector, Neutral Fluid Displacement, 5.7 " (14 cm), REF 7N8370K - Product Usage: intended for single patient use with a vascular access device for the administration of drugs and solutions without needles, thus eliminating the potential for needle-stick injuries during use.

Recall number
Z-1102-2021
Recalling firm
Baxter Healthcare Corporation
Product
Baxter ONE-LINK Non-DEHP Y-Type Microbore Catheter Extension Set, Bonded Needle-free IV Connector, Neutral Fluid Displacement, 5.7 " (14 cm), REF 7N8370K - Product Usage: intended for single patient use with a vascular access device for the administration of drugs and solutions without needles, thus eliminating the potential for needle-stick injuries during use.
Status
Terminated
Initiated
2020-12-21
Posted
not on file
Terminated
2024-07-11
Reason
There were customer reports of separation between the tubing and Male Luer Lock Adaptor.
Root cause
Process design
Action
The firm initiated the recall by email on 12/21/2020. The letter requested the following: 1. Locate and remove the affected product from consignee's facility. The product code and lot number can be found on the individual product or shipping carton. 2. Contact Baxter Healthcare Center for Service to arrange for return and credit. 3. Confirm the receipt of the letter by completing the enclosed Baxter customer reply form and return it to Baxter via email
Quantity
66,800 units
Distribution
Worldwide distribution - US Nationwide distribution in the states of SC, OH, MI, CA, TX, and the countries of UK, Ireland.
Lot, serial or model codes
UDI: 00085412613505, Lot Numbers: DR18E01023, exp. 5/2/2023; DR18E07053, exp. 5/8/2023; DR19G08023, exp. 7/8/2024; DR19H07015, exp. 8/7/2024; DR19L09030, exp. 12/10/2024; DR20B04020, exp. 2/5/2025
510(k) numbers
["K132734"]
PMA numbers
not on file
Product code
FPA
Firm FEI
1417572
Firm city
Deerfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.