FDA Medical Device Recalls (openFDA)

Philips Medical Systems (Cleveland) Inc: Intellispace Portal (ISP) Philips Medical Systems (Cleveland), Inc., IntelliSpace Portal (ISP) is a multimodality (CT, NM, and MR) thin-client applications server that delivers full diagnostic viewing and clinical applications to the enterprise. IntelliSpace Portal (ISP) is a medical software system that allows multiple users to remotely access IntelliSpace Workspace Portal (ISP) from compatible computers on a network. The system allows networking, selection, processing and filming of multimodality DICOM images. Both the client and server software are only for use with off the shelf hardware technology that meets defined minimum specifications. trained physicians may use the images as a basis for diagnosis upon ensuring that monitor quality, ambient light conditions and image compression ratios are consistent with clinical application.

Recall number
Z-1102-2013
Recalling firm
Philips Medical Systems (Cleveland) Inc
Product
Intellispace Portal (ISP) Philips Medical Systems (Cleveland), Inc., IntelliSpace Portal (ISP) is a multimodality (CT, NM, and MR) thin-client applications server that delivers full diagnostic viewing and clinical applications to the enterprise. IntelliSpace Portal (ISP) is a medical software system that allows multiple users to remotely access IntelliSpace Workspace Portal (ISP) from compatible computers on a network. The system allows networking, selection, processing and filming of multimodality DICOM images. Both the client and server software are only for use with off the shelf hardware technology that meets defined minimum specifications. trained physicians may use the images as a basis for diagnosis upon ensuring that monitor quality, ambient light conditions and image compression ratios are consistent with clinical application.
Status
Terminated
Initiated
2013-02-12
Posted
2013-04-12
Terminated
2013-11-04
Reason
Philips was notified of a problem using the IntelliSpace Portal with software version 4.0.2 with the URL PACS integration. Philips was informed that when loading a study to an analysis application while using URL PACS integration configuration, after another study was opened by a review application, the Analysis monitor is updated with the study of the current patient, but the Review monitor conti
Root cause
Software design
Action
Philips sent a Urgent Medical Device Correction letter dated February 12, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. They advised their customers that when reviewing two disparate image datasets, not to only be focused on the actual image content but to also be alerted to the patient's study name and/or ID prior to interpretation. They recommend to manually exit the Review and Analysis applications before loading a new study to Intellispace Portal v4.0.2 from the PACS. For North America and Canada contact the Customer Care Solutions Center at (1-800) 722-9377.
Quantity
313 units
Distribution
Worldwide Distribution - (USA) Nationwide Distribution including the states of AL, AR, AZ, CA, FL, GA, IA, IN, LA, MI, MN, MO, NC, NJ, NY, OH, OK, OR, PA, TN, TX, WA, WI & WV. and the countries of Australia, Austria, Belgium, Canada, China, Czech Republic, Denmark, Egypt, Finland, France, France, Germany, Greece, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Latvia, Lebanon, Monaco, Netherlands, Norway, Poland, Portugal, Russia, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand & United Kingdom
Lot, serial or model codes
Model #881001, Serial #s: 79916, 79932, 79933, 79935, 79936, 79937, 79939, 79940, 79943, 80071, 80914, 80916, 80925, 80926, 80928, 80932, 80939, 80963, 80965, 80977, 80979, 80994, 80995, 81004, 81006, 81008, 81010, 81026, 81027, 81028, 81034, 81035, 81036, 81039, 81063, 81065, 81076, 81079, 81081, 81085, 81089, 81097, 81098, 81108, 81110, 81117, 81119, 81127, 81130, 81132, 81133, 81150, 81153, 81154, 81160, 81161, 81163, 81165, 81169, 81172, 81173, 81174, 81175, 81178, 81179, 81182, 81183, 81184, 81187, 81189, 81190, 81192, 81194, 81196, 81200, 82001, 82002, 82003, 82004, 82006, 82007, 82008, 82009, 82010, 82011, 82012, 82013, 82014, 82015, 82016, 82017, 82018, 82019, 82020, 82021, 82022, 82023, 82024, 82025, 82026, 82027, 82028, 82029, 82030, 82031, 82032, 82033, 82034, 82035, 82036, 82037, 82038, 82050, 82051, 82052, 82053, 82054, 82055, 82056, 82057, 82058, 82059, 82060, 82061, 82062, 82063, 82064, 82065, 82066, 82067, 82068, 82069, 82070, 82071, 82072, 82073, 82074, 82084, 82086, 82087, 82088, 82089, 82090, 82091, 82092, 82095, 82097, 82098, 82099, 82100, 82101, 82102, 82103, 82104, 82105, 82106, 82107, 82108, 82109, 82110, 82112, 82116, 82118, 82120, 82121, 82123, 82125, 82126, 82128, 82129, 82130, 82131, 82132, 82133, 82138, 82142, 82143, 82144, 82146, 82148, 82201, 82202, 82203, 82236, 82238, 82240, 82241, 82250, 82251, 85005, 85006, 85007, 85008, 85009, 85010, 85011, 85014, 85025, 85026, 85027, 85028, 85029, 85031, 85032, 85033, 85040, 85041, 85042, 85043, 85044, 85046, 85047, 85049, 85050, 85066, 85067, 85068, 85069, 85080, 85081, 85082, 85083, 85084, 85086, 85088, 85090, 85096, 85103, 85104, 85105, 85106, 85107, 85108, 85110, 85117, 85141, 85143, 85144, 85145, 85149, 85164, 85165, 85167, 85168, 85169, 85170, 85171, 85172, 85174, 85175, 85176, 85177, 85178, 85179, 85180, 85213, 85214, 85215, 85216, 85217, 85218, 85219, 85238, 85239, 85243, 85244, 85245, 85246, 85247, 85248, 85249, 85250, 85252, 85253, 85254, 85255, 85257, 85260, 85296, 85299, 85300, 85302, 85305, 85306, 85307, 85308, 85309, 85312, 85315, 85316, 85317, 85319, 85320, 85322, 85329, 85330, 85332, 85333, 85334, 85335, 85337, 85338, 85355, 85373, 85475, 95001, 95007, 95008, 95009, 820001, 822002, 822003, 822004, 822005, 822009, 822017, 850004, 850015, 920003 & 950001.
510(k) numbers
not on file
PMA numbers
not on file
Product code
JAK
Firm FEI
1525965
Firm city
Cleveland
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register