FDA Medical Device Recalls (openFDA)
Boston Scientific: Boston Scientific Wiseguide Guide Catheter 7F, .076 '', Femoral Approach, Manufactured for Boston Scientific Scimed, Inc., One Scimed Place, Maple Grove, MN USA 55311. Nonpyrogenic.
- Recall number
- Z-1102-06
- Recalling firm
- Boston Scientific
- Product
- Boston Scientific Wiseguide Guide Catheter 7F, .076 '', Femoral Approach, Manufactured for Boston Scientific Scimed, Inc., One Scimed Place, Maple Grove, MN USA 55311. Nonpyrogenic.
- Status
- Terminated
- Initiated
- 2006-05-22
- Posted
- 2006-06-10
- Terminated
- 2006-12-30
- Reason
- Gaps were identified between polymer segments on the 7F Wiseguide Guide Catheter in which the PTFE was not properly laminated to the braid of the catheter. This delamination of the PTFE from the braid could compromise the device delivery.
- Root cause
- Other
- Action
- Affected customers were mailed an Urgent Medical Device Recall or Correction - Immediate Action Required letter, dated 05/22/2006. The letter describes the issue, states to immediately discontinue use of and segregate recalled product, provides instructions to ship product back to Boston Scientific and asks for the Reply Verification Tracking Form to be faxed to Boston Scientific in acknowledgement.
- Quantity
- 75
- Distribution
- Nationwide- States of AZ, CA, CO, FL, HI, KS, MI, NV, OK, TX, VA.
- Lot, serial or model codes
- Catalog number 19500-448, Lot # 655713; Catalog number 19500-498, Lot # 655721; Catalog number 19500-08, Lot # 657310; Catalog number 19500-09, Lot # 657311; Catalog number 19500-132, Lot # 655680; Catalog number 19500-27, Lot # 655695;
- 510(k) numbers
- ["K974684"]
- PMA numbers
- not on file
- Product code
- DQY
- Firm FEI
- 3002095335
- Firm city
- Maple Grove
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28