Skip to the content

FDA Medical Device Recalls (openFDA)

Boston Scientific: Boston Scientific Wiseguide Guide Catheter 7F, .076 '', Femoral Approach, Manufactured for Boston Scientific Scimed, Inc., One Scimed Place, Maple Grove, MN USA 55311. Nonpyrogenic.

Recall number
Z-1102-06
Recalling firm
Boston Scientific
Product
Boston Scientific Wiseguide Guide Catheter 7F, .076 '', Femoral Approach, Manufactured for Boston Scientific Scimed, Inc., One Scimed Place, Maple Grove, MN USA 55311. Nonpyrogenic.
Status
Terminated
Initiated
2006-05-22
Posted
2006-06-10
Terminated
2006-12-30
Reason
Gaps were identified between polymer segments on the 7F Wiseguide Guide Catheter in which the PTFE was not properly laminated to the braid of the catheter. This delamination of the PTFE from the braid could compromise the device delivery.
Root cause
Other
Action
Affected customers were mailed an Urgent Medical Device Recall or Correction - Immediate Action Required letter, dated 05/22/2006. The letter describes the issue, states to immediately discontinue use of and segregate recalled product, provides instructions to ship product back to Boston Scientific and asks for the Reply Verification Tracking Form to be faxed to Boston Scientific in acknowledgement.
Quantity
75
Distribution
Nationwide- States of AZ, CA, CO, FL, HI, KS, MI, NV, OK, TX, VA.
Lot, serial or model codes
Catalog number 19500-448, Lot # 655713; Catalog number 19500-498, Lot # 655721; Catalog number 19500-08, Lot # 657310; Catalog number 19500-09, Lot # 657311; Catalog number 19500-132, Lot # 655680; Catalog number 19500-27, Lot # 655695;
510(k) numbers
["K974684"]
PMA numbers
not on file
Product code
DQY
Firm FEI
3002095335
Firm city
Maple Grove
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register