FDA Medical Device Recalls (openFDA)
GE Healthcare, LLC: GE Healthcare Precision 600FP, Classical R&F (radiographic and fluoroscopic) System - Product Usage: Intended for use by a qualified/trained doctor or technologist on both adult and pediatric subjects taking diagnostic radiographic and fluoroscopic exposures of the whole body, skull, spinal column, chest, abdomen, extremities and other body parts.
- Recall number
- Z-1101-2021
- Recalling firm
- GE Healthcare, LLC
- Product
- GE Healthcare Precision 600FP, Classical R&F (radiographic and fluoroscopic) System - Product Usage: Intended for use by a qualified/trained doctor or technologist on both adult and pediatric subjects taking diagnostic radiographic and fluoroscopic exposures of the whole body, skull, spinal column, chest, abdomen, extremities and other body parts.
- Status
- Terminated
- Initiated
- 2020-12-10
- Posted
- not on file
- Terminated
- 2022-11-18
- Reason
- GE Healthcare has become aware of a potential safety issue associated with the improper installation of the Precision 600FP monitor suspension. A partial detachment of the monitor suspension could result in potential injury to a patient or operator.
- Root cause
- Other
- Action
- The firm issued an urgent medical device correction notice by letter on 12/10/2020. The letter states: "You can continue to use your Precision 600 FP monitor suspension under the following condition. If you observe any abnormal movement or misalignment of the monitor suspension in day to day performance controls, discontinue use and contact your GE Healthcare Service Representative immediately." Customers with questions or concerns may contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
- Quantity
- 113 units
- Distribution
- US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, FL, GA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, MT, NC, NJ, NM, NV, NY, OH, PA, SC, TN, TX, UT, VA, WA, WI, WV
- Lot, serial or model codes
- Serial Numbers: G1D1842032, G1D1842035, G1C17X2022, G1D1852039, Not Available*, G1D1922084, G1B16X2002, G1D1972106, G1D18X2058, G1D18Y2066, G1B1912074, G1B1772014, G1C1782016, G1D1842033, G1D18Z2071, G1D1962102, G1D1962104, G1D1862041, G1D18Y2068, G1D1972107, G1D18Z2069, Not Available*, G1D19X2118, G1D1982112, Not Available*, G1D1862043, G1D18Y2065, Not Available*, G1D1962101, G1C17X2026, G1D1832029, G1D1962105, G1B1772012, G1D1832030, G1B1772012, G1D1852038, Not Available, G1D1842031, G1C17X2023, G1C1782017, G1D1892054, G1D1932088, G1C17X2024, G1D1882049, G1D1882048, G1D1912077, G1D1932089, GID194092, G1923001923, G1793810017, G1D1882052, G1D1922080, G1D1862045, G1D1882051, G1C1792020, G1D1892055, G1D1932090, G1B1742006, G1D18Z2070, G1D1862042, G1D18X2062, G1D18X2061, G1B1762011, G1B1762010, G1D1942093, G1D1862044, G1D1882050, G1B1912079, G1D1992113, G1B1762009, G1D1852036, G1D1992117, G1D1842034, G1D1922083, G1D1952097, G1D18Z2073, G1D18X2060, G1D1912078, G1D1922085, G1D1892057, Not Available*, G1D1932091, G1D1922081, Not Available*, G1D1912075, G1D18Z2072, G1D1852040, G1D1892056, G1C1792021, G1D1982111, G1D1952100, G1B16X2005, G1D1952096, G1D1962103, G1D1922082, G1D1972108, Not Available*, G1D1932086, G1B16X2004, G1D1852037, G1D18X2059, G1D18Y2064, G1D18Y2063, G1B1942095, G1D1932087, G1D1913076, G1D18Y2067, G1D1872046, G1D1832028, G1D1992114, G1C1792018, G1C17X2025, G1B16X2003 not available* - to be supplied
- 510(k) numbers
- ["K133553"]
- PMA numbers
- not on file
- Product code
- JAA
- Firm FEI
- 2126677
- Firm city
- Waukesha
- Firm state
- WI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28