Skip to the content

FDA Medical Device Recalls (openFDA)

Clarity Medical Systems Inc: Retcam3 & Retcam Shuttle with version 6.0 software. Clarity Medical Systems, Inc. 5775 W. Las Positas Blvd, Suite 200 Pleasanton, CA 94588 General ophthalmic imaging including retinal, corneal and external imaging. Photodocumentation of pediatric ocular diseases including retinopathy or prematurity (ROP). Screening of Type-2 re-threshold retinopathy of prematurity or treatment requiring ROP.

Recall number
Z-1099-2012
Recalling firm
Clarity Medical Systems Inc
Product
Retcam3 & Retcam Shuttle with version 6.0 software. Clarity Medical Systems, Inc. 5775 W. Las Positas Blvd, Suite 200 Pleasanton, CA 94588 General ophthalmic imaging including retinal, corneal and external imaging. Photodocumentation of pediatric ocular diseases including retinopathy or prematurity (ROP). Screening of Type-2 re-threshold retinopathy of prematurity or treatment requiring ROP.
Status
Terminated
Initiated
2012-02-02
Posted
2012-02-24
Terminated
2012-06-07
Reason
Still images extracted from video recordings may be mislabeled as to right and left eye.
Root cause
Software design
Action
Clarity Medical Systems, Inc. sent a Safety Alert letter dated February 7, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The customer was instructed the following : 1. Do not use the video mode until the software has been updated. 2. Install the software update using the enclosed CD. The instructions for installation are also enclosed. This software update will correct the anomaly. 3. If you have extracted still images from video recordings, review all potentially affected patient files to determine if any right eye images are mislabeled as left eye images and/or left eye images are mislabeled as right eye images. If any images are mislabeled, make the appropriate corrections in the annotation section of the affected file. Consider whether there is any diagnostic impact or need to re-irnage the patient. We apologize for the inconvenience. If you have questions or concerns, please contact Clarity at 800-215-6005.
Quantity
33 RetCam Shuttles and 20 RetCam 3 - total 53.
Distribution
Worldwide Distribution -- USA (nationwide) including the states of TN, MO, TX, FL, NJ and Puerto Rico and countries of Italy, Portugal, Czech Republic, Thailand, Chile, Oman, Hong Kong, Russia, Philippines, Sweden, Finland, Ecuador, Australia and Canada.
Lot, serial or model codes
Model number RC3, part number 21-100500; Model number RCS, part number 20-000300.
510(k) numbers
["K081858","K090326","K102859"]
PMA numbers
not on file
Product code
HKI
Firm FEI
3004014430
Firm city
Pleasanton
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register