Ethicon, LLC: ETHIBOND EXCEL Polyester Suture, REF: 6664H, X425H, X872H, EH7491H. Not sold in US

Recall number
Z-1098-2025
Recalling firm
Ethicon, LLC
Product
ETHIBOND EXCEL Polyester Suture, REF: 6664H, X425H, X872H, EH7491H. Not sold in US
Status
Open, Classified
Initiated
2024-12-20
Posted
2025-02-04
Terminated
not on file
Reason
Sutures may have an open seal on the primary packaging, due to a manufacturing issue on a specific machine, and if issue is undetected could lead to breach in sterility, could introduce pathogens to the patient and cause infection.
Root cause
Process control
Action
On 12/20/2024, Johnson & Johnson Med Tech mailed recall notices to Operating Room Supervisors, Materials Management Personnel, Chief of Surgery, and distributors who were asked to do the following: 1) Examine your inventory immediately to determine if you have product subject to this recall on hand and quarantine such product(s). 2) Communicate the issue to relevant operating room or materials management personnel, or anyone else in your facility who needs to be informed. If any product subject to this recall has been forwarded to another facility, contact that facility to arrange return. Please consider including a copy of this recall letter when communicating. 3) Contact Sedgwick at 888-548-8526 to arrange affected product return. 4) Keep this notice visibly posted for awareness until all product subject to this recall has been returned to Sedgwick. 5) Complete and return the Business Reply Form via email to ethicon3703@sedgwick.com If you have additional questions regarding this communication contact the firm's Resource Department at 1-877-ETHICON (1-877-384-4266) open Monday through Friday, 8:00 AM to 5:00 PM ET.
Quantity
48
Distribution
Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, LA MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI and the countries of AT, AU, BE, CA, CH, CN, CZ, DE, DK, FI, FR, GB, HU, IE, JP, LU, NL, NO, PL, SE, SG, ZA.
Lot, serial or model codes
REF/GTIN/Lot(expiration): W8003T/10705031231658/103GT2(8/31/2029)
510(k) numbers
not on file
PMA numbers
not on file
Product code
GAW
Firm FEI
2648650
Firm city
San Lorenzo
Firm state
not on file
Firm country
PR

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register