Skip to the content

FDA Medical Device Recalls (openFDA)

Soft Computer Consultants, Inc.: Softbank software Product Usage: Decision support software for transfusion service. It keeps track of inventory from outside sources, multi-site inventory control, records of testing of units, and allows for record keeping for transfusion preparation.

Recall number
Z-1098-2017
Recalling firm
Soft Computer Consultants, Inc.
Product
Softbank software Product Usage: Decision support software for transfusion service. It keeps track of inventory from outside sources, multi-site inventory control, records of testing of units, and allows for record keeping for transfusion preparation.
Status
Terminated
Initiated
2016-10-06
Posted
not on file
Terminated
2019-08-21
Reason
Software error. Potential for incorrect results
Root cause
Software design (manufacturing process)
Action
Consignees were sent a recall notification, dated 10/06/2016, providing instructions for software correction. The firm indicates that the issue will be resolved with a mandatory hot fix. Hot fix will be released when its effectiveness is confirmed. Consignees are instructed to acknowledge receipt of the recall notification and grant permission for the recalling firm to load the hot fix when available. The firm can be contacted by calling Technical Support at 1-800-763-8522. Until the hot fix is available, consignees can still operate the software by exiting (Esc-Quit) the Test Verify option after each save changes (F12-Accept).
Quantity
189 total
Distribution
Worldwide Distribution - US Nationwide Distribution in the states of AL, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NH, NJ, NV, NY, OH, OK, OR, PA, RI, TN, TX, VA, WA, WI, WV, and the countries of Hawaii and Puerto Rico Canada & Jamaica
Lot, serial or model codes
Version: 35.5.2.0, 25.5.3.0, 25.5.3.1
510(k) numbers
not on file
PMA numbers
not on file
Product code
NSX
Firm FEI
1000307651
Firm city
Clearwater
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register