Skip to the content

FDA Medical Device Recalls (openFDA)

Deerfield Imaging: Intra operative MRI Systems: Neuro II-SE , Neuro III-SV, iMRI 1.5T A, and iMRI 3T S. Indicated for use as a magnetic resonance diagnostic device (MRDD).

Recall number
Z-1098-2016
Recalling firm
Deerfield Imaging
Product
Intra operative MRI Systems: Neuro II-SE , Neuro III-SV, iMRI 1.5T A, and iMRI 3T S. Indicated for use as a magnetic resonance diagnostic device (MRDD).
Status
Terminated
Initiated
2016-02-10
Posted
2016-03-11
Terminated
2017-04-05
Reason
Emergency helium venting lines (i.e. quench lines) may not have been properly installed.
Root cause
Device Design
Action
Consignees were sent a IMRIS "Urgent Field Safety Notice" letter dated February 10, 2016. The letter described the problem and the product involved in the recall. The firm also stated that it is important that customers review the portion of the quench line vent described and ensure that it is installed to the specifications supplied as part of the Planning Guide. Customers were reminded that it is important to have an emergency plan in place in the event Helium gas escapes into the magnet room or any other area. For questions customers can contact their local customer support hotline or +1-866-475-0525.
Quantity
35 USA, 14 OUS
Distribution
Worldwide Distribution -- US, including the states of CA, CT, FL, GA, IL, IN, MD, MA, MN, MO, NH, NC, OH, OR, PA, TX, UT, VA, and WI; and ,countries of CHINA, CANADA, AUSTRALIA, FRANCE, JAPAN, and GERMANY.
Lot, serial or model codes
70036, 10076-21, M00004, 10073-01, 23529, 40481, M00152, M00005, 30445, M00169, 060014, 10043-00, 30823, 10004680, 10050-01, 40662, 30935, 31084, 31203, 10002221, 40510, 40613, 30719, 31046, 30910, 30832, 30623, 31213, 10004152, 10072-01, 31250, 40815, 31049, 30068, 40412, M0038, M00106, M00367, 10035-01, 40475, 40495, 10007498, 10005946, 40263, 40688, 10058-00, 40619, 31085, and 10005639.
510(k) numbers
["K061916","K071099","K083137","K133692"]
PMA numbers
not on file
Product code
LNH
Firm FEI
3010326005
Firm city
Minnetonka
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register