FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: Iconos R200 C20, Iconos R200 T20, Iconos R100, Iconos MD, Iconos RHD, Luminos RF Classic, Luminos Fusion or Luminos Select. Angiographic x-ray system.

Recall number
Z-1097-2013
Recalling firm
Siemens Medical Solutions USA, Inc
Product
Iconos R200 C20, Iconos R200 T20, Iconos R100, Iconos MD, Iconos RHD, Luminos RF Classic, Luminos Fusion or Luminos Select. Angiographic x-ray system.
Status
Terminated
Initiated
2013-01-15
Posted
2013-04-11
Terminated
2013-04-23
Reason
Siemens has become aware of a potential malfunction and possible hazard to patients and staff when using the Iconos R200 C20, Iconos R200 T20, Iconos R100, Iconos MD, Iconos RHD, Luminos RF Classic, Luminos Fusion or Luminos Select. This malfunction is possible in cases where the table vertical lift movement is obstructed by a solid object which could potentially result in collision. This can re
Root cause
Process design
Action
Siemens sent a Customer Safety Advisory letter dated January 15th, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The SIEMENS service representative will investigate if parts need to be exchanged in your product, an if so, which parts need to be exchanged to restore the system safety and function. Please kindly confirm receipt of the Field Safety Notice by sending back the attached form. For further questions please call (610) 448-3237.
Quantity
51
Distribution
Nationwide Distribution including the states of CA, FL, GA, IA, IN, KS, LA, ME, MI, MO, NC, NH, NJ, NY, NV, OH, PA, TN, TX, VA and WA.
Lot, serial or model codes
Model Number 5902767 with serial numbers 2217, 2038, 1228, 1007, 1219, 2156, 2264, 2190, 2132, 2213, 1484, 2166, 2259, 2068, 1610, 1601, 1413, 2236, 2153, 1444, 1477, 1632, 2037, 2180, 1663, and 1612; Model Number 5902783 with serial numbers 1003, 1401, 1402, and 1405; Model Number 7152478 with serial numbers 2580, 2524, 2590, 2536, 2563, and 2534; Model Number 10093962 with serial numbers 5837, 5120, 5523, 5910, 5507, 6141, 4112, 6274, 4506, 5900, 4121, 6245, 6242, 4509, and 6113.
510(k) numbers
["K992660"]
PMA numbers
not on file
Product code
JAA
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register