Skip to the content

FDA Medical Device Recalls (openFDA)

Abbott Diagnostic International, Ltd.: Architect AUSAB Calibrators IVD - list number 1L82-01 has six (6) bottles with the following companents: 1) Calibrator A and 2) Calibrators B thru F. in vitro diagnostic.

Recall number
Z-1097-2007
Recalling firm
Abbott Diagnostic International, Ltd.
Product
Architect AUSAB Calibrators IVD - list number 1L82-01 has six (6) bottles with the following companents: 1) Calibrator A and 2) Calibrators B thru F. in vitro diagnostic.
Status
Terminated
Initiated
2007-05-14
Posted
2007-07-24
Terminated
2011-06-10
Reason
During an internal study, the Architect AUSAB assay was found to read preparations of the WHO standard up to 53% higher than the WHO assigned value. This has the potential to cause elevated results and may impact the interpretation for a small percentage of results with grayzone or low-level reactive concentrations and higher proficiency sample results.
Root cause
Other
Action
All ARCHITECT AUSAB U.S. customers will be contacted. New customers will be contacted through a Product Information Letter that will be included in all ARCHITECT AUSAB Reagent Kits until assay is restandardized. All ARCHITECT AUSAB U.S. customers will be contacted. The following communication types were used: Product Correction Letter Customer Reply Form > Product Information Letter to be included in new kits to be distributed. Letters were sent to customers by Federal Express Priority mail. The Customers were instructed to: Please follow their laboratory procedures for evaluation of these results Disregard the automated interpretation of results for low-level reactive samples. A customer reply form will accompany the Product Correction letter. Effectiveness Check Depth: 100% of customers receiving the Product Correction letter. The start date of communication was May 14,2007.
Quantity
247
Distribution
Nationwide and Puerto Rico
Lot, serial or model codes
LIST # 01L82-01 Lots # 50337Q100 Exp Date 08/21/2007, 48265Q100 Exp Date 06/13/2007, 45555Q100 Exp Date 03/14/2007, 44594Q100 Exp date 02/28/2007, and 42585Q100 Exp Date 01/01/2007
510(k) numbers
not on file
PMA numbers
["P050051"]
Product code
LOM
Firm FEI
2623532
Firm city
Barceloneta
Firm state
PR
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register