Skip to the content

FDA Medical Device Recalls (openFDA)

Microbiologics Inc: Vaginal Verification Panel Ref 8208 Lot 8208-11

Recall number
Z-1096-2020
Recalling firm
Microbiologics Inc
Product
Vaginal Verification Panel Ref 8208 Lot 8208-11
Status
Terminated
Initiated
2020-01-03
Posted
not on file
Terminated
2020-09-16
Reason
Product 8208 Vaginal Verification Panel is a 6 pool kit and has contaminant Candida krusei in Pool 2 (subcomponent 6063). Candida krusei is an organism intended to be contained in Pool 6 of the kit. This should have no impact on patient test results as long as Gardnerella vaginalis was properly detected when using Pool 2 and/or users are strictly using this product for instrument verification/validation.
Root cause
Process control
Action
Letters will be sent via email to all consignees and end users regarding the correction. Users are to review the instructions, determine how the information affects their usage, and upon determination either use with new information or discard the product.
Quantity
9
Distribution
Distributed in the US to 7 states: CA, KS, ME, NC, NY, OR, and UT. No Canada, Mexico or VA.
Lot, serial or model codes
Lot 8208-11
510(k) numbers
not on file
PMA numbers
not on file
Product code
OHQ
Firm FEI
2150138
Firm city
Saint Cloud
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register