Hitachi Medical Systems America Inc: Hitachi Oasis MRI System - C-Spine Coil
- Recall number
- Z-1096-2017
- Recalling firm
- Hitachi Medical Systems America Inc
- Product
- Hitachi Oasis MRI System - C-Spine Coil
- Status
- Terminated
- Initiated
- 2017-01-06
- Posted
- not on file
- Terminated
- 2018-11-08
- Reason
- The customer indicated that the coil balun was hot to the touch when removing the coil from the patient table after an exam.
- Root cause
- Component design/selection
- Action
- Hitachi sent an Urgent : Device Correction Notice dated January 3, 2017, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The firm plans to visit each site and replace the cable balun and also confirm receipt of the notification letter. The firm will quarantine returned baluns in their warehouse until they are reworked. For further questions regarding this recall please call (330) 425-1313 ext. 3720.
- Quantity
- 180 units
- Distribution
- USA (nationwide) distribution. The product was distributed to the following states: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NH, NJ, NM, NY, OH, OK, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WY. There are three (3) government accounts for this recall: VA Medical Center/Lyster Army Health Clinic, Parma VA Medical Center, Outpatient Clinic. There are no Canada accounts for this recall. There are no Mexico accounts for this recall.
- Lot, serial or model codes
- M001 to M028, M029 to M083, M085 to M090, M092 to M104, M106, M110 to M111, M113 to M114, M116 to M123, M125, M128 to M131, M133, M139 to M141, M143 to M147, M150, M152, M154, M159, M161, M163, M166, M169 to M176, M179, M182, M183, M185, M186, M190, M194 to 196, M201, M207 to M238
- 510(k) numbers
- ["K072279"]
- PMA numbers
- not on file
- Product code
- LNH
- Firm FEI
- 1528028
- Firm city
- Twinsburg
- Firm state
- OH
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28