FDA Medical Device Recalls (openFDA)

Hologic, Inc.: The Rapid fFN Specimen Collection Tube (PN: 71550-001) is a component of the Rapid fFN Specimen Collection kit (PN: 71738-001). The Collection Tube (PN: 71550-001) is manufactured by a third-party supplier, MML Diagnostics Packaging, and includes the transport tube, cap, and extraction buffer. Tubes are packaged into PN: 71738-00 at Hologic prior to kit release.

Recall number
Z-1095-2025
Recalling firm
Hologic, Inc.
Product
The Rapid fFN Specimen Collection Tube (PN: 71550-001) is a component of the Rapid fFN Specimen Collection kit (PN: 71738-001). The Collection Tube (PN: 71550-001) is manufactured by a third-party supplier, MML Diagnostics Packaging, and includes the transport tube, cap, and extraction buffer. Tubes are packaged into PN: 71738-00 at Hologic prior to kit release.
Status
Open, Classified
Initiated
2025-01-08
Posted
2025-02-04
Terminated
not on file
Reason
Three lots of Rapid fFN Specimen Collection kit (901164, 901165, and 904850) contain Rapid fFN Specimen Collection Tubes which may be empty, underfilled, or overfilled with extraction buffer. The transport tubes are used with the Fetal Fibronectin Enzyme Immunoassay and Rapid fFN for the TLiIQ System which aids in assessing the risk of preterm delivery. The use of empty transport tubes or transport tubes containing insufficient extraction buffer volume may result in invalid results and/or delays in obtaining results. Additionally, the use of underfilled tubes may result in false positive results and the use of overfilled tubes may result in false negative results.
Root cause
Nonconforming Material/Component
Action
Hologic issued a "Urgent: Medical Device Recall Notification on January 8, 2025 via FedEx. Hologic asked consignees to take the following actions: 1. Please discontinue use of the Rapid fFN Specimen Collection Kit Catalog Number: 71738-001; 2. Please segregate and count how many of the affected Rapid fFN Specimen Collection Kit Catalog Number: 71738-001; 3. Please fill out the Customer Response Form (CRF); 4. Hologic will process and ship the replacement kits as listed on the completed CRF once it has been received; 5. For tests already administered from the affected product lots and if the patient has not given birth, it is at the discretion of the healthcare providers to notify patients who were managed based on a number of clinical decision points including but not exclusive to fetal fibronectin test results. If the patient has given birth, no further action is required.; 6. If you have any questions or concerns regarding this recall notification, please contact Hologic Technical Support
Quantity
1351 units
Distribution
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, Guam and the countries of Canada and Japan.
Lot, serial or model codes
Lot # 901164, 901165, 904850/ UDI: 25420045514765
510(k) numbers
["K900225"]
PMA numbers
not on file
Product code
LIO
Firm FEI
2024800
Firm city
San Diego
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register