MEDLINE INDUSTRIES, LP - Northfield: Sterile Procedural Trays, labeled as the following: a. BASIC EYE PACK-LF b. VITRECTOMY PACK-LF c. EYE PACK-EAST-LF d. OSC OCULAR PLASTIC PACK-LF e. EYE PACK f. OCULOPLASTIC PACK g. EYE PLASTIC PACK h. PLASTIC EYE PACK i. DR H BLEPH PACK j. BLEPH PACK k. OPHTHALMOLOGY PACK l. OCULOPLASTIC PACK m. BLEPH PACK n. EYE PACK o. OCULAR PLASTIC PACK p. OCULOPLASTIC PACK
- Recall number
- Z-1095-2023
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Product
- Sterile Procedural Trays, labeled as the following: a. BASIC EYE PACK-LF b. VITRECTOMY PACK-LF c. EYE PACK-EAST-LF d. OSC OCULAR PLASTIC PACK-LF e. EYE PACK f. OCULOPLASTIC PACK g. EYE PLASTIC PACK h. PLASTIC EYE PACK i. DR H BLEPH PACK j. BLEPH PACK k. OPHTHALMOLOGY PACK l. OCULOPLASTIC PACK m. BLEPH PACK n. EYE PACK o. OCULAR PLASTIC PACK p. OCULOPLASTIC PACK
- Status
- Open, Classified
- Initiated
- 2022-11-17
- Posted
- 2023-02-06
- Terminated
- not on file
- Reason
- Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
- Root cause
- Device Design
- Action
- Medline Industries, LP notified consignees via email and first-class mail on 11/15/2022. Consignees were instructed to check stock for affected product and immediately quarantine. For any affected units, customers were instructed to over-label the sterile kits with a warning label indicating "Warning To Avoid Injury, Please Handle Blades with Care." No product is to be returned. Customers were also instructed to complete and return the response form and notify customers if further distributed.
- Quantity
- 260 cases (1,200 units)
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, Japan, Mexico, Panama, Turks & Caicos, United Arab Emirates.
- Lot, serial or model codes
- a. BASIC EYE PACK-LF, Item Number: DYNJ0842521C, Case UDI/GTIN: 40193489352383, Unit UDI/GTIN: 10193489352382, Lot Number: 22JDA005,22HDC265; b. VITRECTOMY PACK-LF, Item Number: DYNJ0843936F, Case UDI/GTIN: 40193489352390, Unit UDI/GTIN: 10193489352399, Lot Number: 22JDA822,22IDB984; c. EYE PACK-EAST-LF, Item Number: DYNJ14064, Case UDI/GTIN: 40080196931102, Unit UDI/GTIN: 10080196931101, Lot Number: 22JBC143,22IBF606; d. OSC OCULAR PLASTIC PACK-LF, Item Number: DYNJ31639B, Case UDI/GTIN: 40889942292067, Unit UDI/GTIN: 10889942292066, Lot Number: 22JBG389,22HBY420; e. EYE PACK, Item Number: DYNJ35056A, Case UDI/GTIN: 40889942227564, Unit UDI/GTIN: 10889942227563, Lot Number: 22JBL816; f. OCULOPLASTIC PACK, Item Number: DYNJ46539J, Case UDI/GTIN: 40193489997355, Unit UDI/GTIN: 10193489997354, Lot Number: 22JME546,22JME931; g. EYE PLASTIC PACK, Item Number: DYNJ47859C, Case UDI/GTIN: 40193489916066, Unit UDI/GTIN: 10193489916065, Lot Number: 22JBS972; h. PLASTIC EYE PACK, Item Number: DYNJ49413A, Case UDI/GTIN: 40888277971593, Unit UDI/GTIN: 10888277971592, Lot Number: 22KBA929; i. DR H BLEPH PACK, Item Number: DYNJ54487A, Case UDI/GTIN: 40193489587877, Unit UDI/GTIN: 10193489587876, Lot Number: 22HBG403; j. BLEPH PACK, Item Number: DYNJ62174A, Case UDI/GTIN: 40195327037780, Unit UDI/GTIN: 10195327037789, Lot Number: 22IBT956; k. OPHTHALMOLOGY PACK, Item Number: DYNJ63785, Case UDI/GTIN: 40193489760744, Unit UDI/GTIN: 10193489760743, Lot Number: 22IBN397; l. OCULOPLASTIC PACK, Item Number: DYNJ66365, Case UDI/GTIN: 40193489346627, Unit UDI/GTIN: 10193489346626, Lot Number: 22JBE699; m. BLEPH PACK, Item Number: DYNJ67875, Case UDI/GTIN: 40193489448826, Unit UDI/GTIN: 10193489448825, Lot Number: 22JBY558,22JBY531; n. EYE PACK, Item Number: DYNJ69446, Case UDI/GTIN: 40193489942546, Unit UDI/GTIN: 10193489942545, Lot Number: 22IBS350,22HBN241; o. OCULAR PLASTIC PACK, Item Number: DYNJ82588, Case UDI/GTIN: 40195327198795, Unit UDI/GTIN: 10195327198794, Lot Number: 22HBJ348; p. OCULOPLASTIC PACK, Item Number: DYNJ908778, Case UDI/GTIN: 40195327177660, Unit UDI/GTIN: 10195327177669, Lot Number: 22JBW843
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- OJK
- Firm FEI
- 1417592
- Firm city
- Northfield
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28