Skip to the content

FDA Medical Device Recalls (openFDA)

Dako North America Inc.: Product Name: HER2 CISH pharmDx Kit Catalog/Model number: SK109, lot 20000910 Product Usage: For in vitro diagnostic use. HER2 CISH pharmDx Kit is intended for dual-color chromogenic visualization of signals achieved with directly labeled in situ hybridization probes targeting the HER2 gene and centromeric region of chromosome 17. The kit is designed to quantitatively determine HER2 gene status in formalin-fixed, paraffin-embedded breast cancer tissue specimens. Red and blue chromogenic signals are generated on the same tissue section for evaluation under bright field microscopy. The CISH procedure is automated using Dako Autostainer instruments. HER2 CISH pharmDx Kit is indicated as an aid in the assessment of patients for whom Herceptin (trastuzumab) treatment is being considered. Results from the HER2 CISH pharmDx Kit are intended for use as an adjunct to the clinicopathologic information currently used for estimating prognosis in stage II, node-positive breast cancer patients.

Recall number
Z-1093-2014
Recalling firm
Dako North America Inc.
Product
Product Name: HER2 CISH pharmDx Kit Catalog/Model number: SK109, lot 20000910 Product Usage: For in vitro diagnostic use. HER2 CISH pharmDx Kit is intended for dual-color chromogenic visualization of signals achieved with directly labeled in situ hybridization probes targeting the HER2 gene and centromeric region of chromosome 17. The kit is designed to quantitatively determine HER2 gene status in formalin-fixed, paraffin-embedded breast cancer tissue specimens. Red and blue chromogenic signals are generated on the same tissue section for evaluation under bright field microscopy. The CISH procedure is automated using Dako Autostainer instruments. HER2 CISH pharmDx Kit is indicated as an aid in the assessment of patients for whom Herceptin (trastuzumab) treatment is being considered. Results from the HER2 CISH pharmDx Kit are intended for use as an adjunct to the clinicopathologic information currently used for estimating prognosis in stage II, node-positive breast cancer patients.
Status
Terminated
Initiated
2014-01-21
Posted
2014-02-26
Terminated
2014-06-04
Reason
Dako has initiated a recall of HER2 CISH pharmDxTM Kit because weak red signals have been observed.
Root cause
Device Design
Action
Dako sent a Recall Notification letter dated January 21, 2014to affected customers via mail. The letter identified the affected product, description of the problem, actions to be taken by the user, communication, and Dako contact information. The letter was accompanied by a "Recall Form (Product Recall Form to customers)". For questions call 805-566-5464.
Quantity
49
Distribution
Worldwide Distribution - USA Nationwide and in the country of Chile.
Lot, serial or model codes
Catalog/Model number: SK109, lot 20000910. Shelf Life: 10 months when stored at 2-8 ¿C. Expiration Date: 2014 June 30.
510(k) numbers
not on file
PMA numbers
["P100024"]
Product code
NYQ
Firm FEI
2022180
Firm city
Carpinteria
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register