Skip to the content

FDA Medical Device Recalls (openFDA)

International Technidyne Corp.: Hemochron Whole Blood Coagulation Systems Activated Clotting Time (ACT); Package label states -International Technidyne Corporation Response Celite ACT Tube HRFTCA510; Lot number and expiration date are listed. Each box contains 95 ACT tubes tests. Used for heparin anticoagulation monitoring during cardiopulmonary bypass surgery, percutaneous transluminal coronary angioplasty (PTCA), interventional radiology, extracorporeal membrane oxygenation (ECMO), hemofiltration, hemodialysis and critical care.

Recall number
Z-1093-2011
Recalling firm
International Technidyne Corp.
Product
Hemochron Whole Blood Coagulation Systems Activated Clotting Time (ACT); Package label states -International Technidyne Corporation Response Celite ACT Tube HRFTCA510; Lot number and expiration date are listed. Each box contains 95 ACT tubes tests. Used for heparin anticoagulation monitoring during cardiopulmonary bypass surgery, percutaneous transluminal coronary angioplasty (PTCA), interventional radiology, extracorporeal membrane oxygenation (ECMO), hemofiltration, hemodialysis and critical care.
Status
Terminated
Initiated
2010-11-30
Posted
2011-02-11
Terminated
2012-08-22
Reason
Some Activated Clotting Time (ACT) Tubes may contain a component that does not meet its specification, making it possible for a failure to detect a tube or the presence of a clot.
Root cause
Process control
Action
ITC notified their distributors and end user customers with an Urgent Medical Device Recall letter dated November 29, 2010 through UPS overnight mail. The letter identified the affected product and explained the reason for the recall. Distributors and customers should check their inventory for any affected product. Distributors should immediately discontinue shipment of the affected product and place the product on hold. Then regardless of whether they have any remaining inventory, they should complete the enclosed Distributor Account Tracking Form. The firm will contact them to arrange for any affected product to be returned and will credit them for any unused, returned product. Customers are to stop using the affected product and remove it from their inventory. They should also complete the attached Customer Account Tracking Form and return it to the firm. The firm will contact them to arrange for the return of any affected inventory and will credit the customers for any unused, returned product. If customers believe that they have experienced a problem with the product, they should take any additional actions regarding their records or patients. Questions should be directed to ITC Nexus Dx Technical Support at 800-631-5945 (US) or 732-548-5700 (International) or e-mail techsupport@itcmed.com.
Quantity
265 boxes (95 ACT Tubes per box)
Distribution
Worldwide Distribution
Lot, serial or model codes
Lot Number G0FTE157A, Expiration Date 7/31/2013 Lot Number G0FTE158A, Expiration Date 7/31/2013 Lot Number G0FTE159A, Expiration Date 7/31/2013 Lot Number G0FTE159B, Expiration Date 7/31/2013
510(k) numbers
["K983475"]
PMA numbers
not on file
Product code
KQG
Firm FEI
1000526865
Firm city
Edison
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register