Skip to the content

FDA Medical Device Recalls (openFDA)

C-RAD POSITIONING AB: Brand Name: Catalyst+ Product Name: Catalyst, Catalyst+ Model/Catalog Number: SP001-0026, SP003-0002 Software Version: cSRS module in combination with cMotion module Product Description: GMDN: 40893 Patient positioning device, diagnostic imaging/radiotherapy, laser Component: Not applicable

Recall number
Z-1092-2026
Recalling firm
C-RAD POSITIONING AB
Product
Brand Name: Catalyst+ Product Name: Catalyst, Catalyst+ Model/Catalog Number: SP001-0026, SP003-0002 Software Version: cSRS module in combination with cMotion module Product Description: GMDN: 40893 Patient positioning device, diagnostic imaging/radiotherapy, laser Component: Not applicable
Status
Open, Classified
Initiated
2025-10-09
Posted
2026-01-14
Terminated
not on file
Reason
Due to issues with the system's stereotactic radiosurgery (SRS) treatment cannot be guaranteed for all couch angles and system setups. n certain cases, the system may indicate the patient as correctly positioned at the isocenter, even when the isocenter is positioned outside of the indicated tolerance.
Root cause
Under Investigation by firm
Action
On 10/14/2025, the firm email an Urgent Field Safety Notice to customers informing them that C-RAD has noted a potential issue affecting the cSRS mode of the Catalyst+ HD and Catalyst HD systems. All software version above c4D 5.3.2 are affected by this issue. The issue is that the stated precision of the Catalyst system for stereotactic radiosurgery (SRS) treatments cannot be guaranteed for all couch angles and system setups. In certain cases, the system may indicate the patient as correctly positioned at the isocenter, even when the isocenter is positioned outside of the indicated tolerance. This could in the worst-case result in erroneous dose delivery. This issue affects Catalyst+ HD and Catalyst HD systems, configured for cSRS and cMotion. Customers are instructed to: 1. Ensure that all the staff operating the Catalyst+ HD and Catalyst HD, and the c4D software application are informed of this information. 2. As an immediate action, affected users are advised to not use Catalyst+ HD and Catalyst HD in combination with cSRS and cMotion. C-RAD will release a new version of the software application c4D resolving the issue and/or update IFU/labelling. For questions please contact C-RAD Quality Department as regulatory@c-rad.com
Quantity
189
Distribution
The software module cSRS has been installed in 37 countries Worldwide - U.S. Nationwide distribution in the states of AK, AL, AR, CA, CT, FL, MO, NC, NM, NY, OH, OK, OR, PA, TN, TX, VA, and WA. The countries of Argentina, Australia, Austria, Belarus, Belgium, Benin, Canada, China, Colombia, Egypt, El Salvador, Finland, France, Germany, Greece, Hungary, India, Italy, Japan, Jordan, Mexico, Morocco, Netherlands, New Zealand, Norway, Portugal, Romania, Russia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Ukraine, and Vietnam.
Lot, serial or model codes
Lot Code: Note: The error is not related to any serial numbers or LOT numbers. The error is limited to the software module cSRS in combination with cMotion module
510(k) numbers
K200435
PMA numbers
not on file
Product code
IYE
Firm FEI
3006621300
Firm city
Uppsala
Firm state
not on file
Firm country
Sweden

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this source