FDA Medical Device Recalls (openFDA)
C-RAD POSITIONING AB: Brand Name: Catalyst+ Product Name: Catalyst, Catalyst+ Model/Catalog Number: SP001-0026, SP003-0002 Software Version: cSRS module in combination with cMotion module Product Description: GMDN: 40893 Patient positioning device, diagnostic imaging/radiotherapy, laser Component: Not applicable
- Recall number
- Z-1092-2026
- Recalling firm
- C-RAD POSITIONING AB
- Product
- Brand Name: Catalyst+ Product Name: Catalyst, Catalyst+ Model/Catalog Number: SP001-0026, SP003-0002 Software Version: cSRS module in combination with cMotion module Product Description: GMDN: 40893 Patient positioning device, diagnostic imaging/radiotherapy, laser Component: Not applicable
- Status
- Open, Classified
- Initiated
- 2025-10-09
- Posted
- 2026-01-14
- Terminated
- not on file
- Reason
- Due to issues with the system's stereotactic radiosurgery (SRS) treatment cannot be guaranteed for all couch angles and system setups. n certain cases, the system may indicate the patient as correctly positioned at the isocenter, even when the isocenter is positioned outside of the indicated tolerance.
- Root cause
- Under Investigation by firm
- Action
- On 10/14/2025, the firm email an Urgent Field Safety Notice to customers informing them that C-RAD has noted a potential issue affecting the cSRS mode of the Catalyst+ HD and Catalyst HD systems. All software version above c4D 5.3.2 are affected by this issue. The issue is that the stated precision of the Catalyst system for stereotactic radiosurgery (SRS) treatments cannot be guaranteed for all couch angles and system setups. In certain cases, the system may indicate the patient as correctly positioned at the isocenter, even when the isocenter is positioned outside of the indicated tolerance. This could in the worst-case result in erroneous dose delivery. This issue affects Catalyst+ HD and Catalyst HD systems, configured for cSRS and cMotion. Customers are instructed to: 1. Ensure that all the staff operating the Catalyst+ HD and Catalyst HD, and the c4D software application are informed of this information. 2. As an immediate action, affected users are advised to not use Catalyst+ HD and Catalyst HD in combination with cSRS and cMotion. C-RAD will release a new version of the software application c4D resolving the issue and/or update IFU/labelling. For questions please contact C-RAD Quality Department as regulatory@c-rad.com
- Quantity
- 189
- Distribution
- The software module cSRS has been installed in 37 countries Worldwide - U.S. Nationwide distribution in the states of AK, AL, AR, CA, CT, FL, MO, NC, NM, NY, OH, OK, OR, PA, TN, TX, VA, and WA. The countries of Argentina, Australia, Austria, Belarus, Belgium, Benin, Canada, China, Colombia, Egypt, El Salvador, Finland, France, Germany, Greece, Hungary, India, Italy, Japan, Jordan, Mexico, Morocco, Netherlands, New Zealand, Norway, Portugal, Romania, Russia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Ukraine, and Vietnam.
- Lot, serial or model codes
- Lot Code: Note: The error is not related to any serial numbers or LOT numbers. The error is limited to the software module cSRS in combination with cMotion module
- 510(k) numbers
- K200435
- PMA numbers
- not on file
- Product code
- IYE
- Firm FEI
- 3006621300
- Firm city
- Uppsala
- Firm state
- not on file
- Firm country
- Sweden
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28