Skip to the content

FDA Medical Device Recalls (openFDA)

GE Medical Systems, LLC: GE Innova 3100 / 3100 IQ, Cardiovascular Imaging System, GE Healthcare, Waukesha, WI 53188 (The system consists of an a monoplane positioner, a vascular or cardiac table, an X-ray system and a digital detector) Indicated for use in diagnostic and interventional angiographic procedures of human anatomy. Intended to replace image intensifier fluoroscopic systems in all diagnostic or interventional procedures. Not intended for mammography applications.

Recall number
Z-1092-2009
Recalling firm
GE Medical Systems, LLC
Product
GE Innova 3100 / 3100 IQ, Cardiovascular Imaging System, GE Healthcare, Waukesha, WI 53188 (The system consists of an a monoplane positioner, a vascular or cardiac table, an X-ray system and a digital detector) Indicated for use in diagnostic and interventional angiographic procedures of human anatomy. Intended to replace image intensifier fluoroscopic systems in all diagnostic or interventional procedures. Not intended for mammography applications.
Status
Terminated
Initiated
2008-03-06
Posted
2009-03-12
Terminated
2014-09-04
Reason
GE Healthcare reported an issue with the Exposure Switch Redundancy system with Innova Systems 3100 & 4100, in which this error stays on, even after the pedal is released, inhibiting further x-rays, and a system reset is required to recover.
Root cause
Software design
Action
On site visits were performed from 3/6/08 - 12/19/08 to update software on all 5 units.
Quantity
4
Distribution
Worldwide Distribution -- USA including states of MD, TX, and VA and country of China.
Lot, serial or model codes
82416120006, 4103323100, 82416120005, and 214324INNOVA2
510(k) numbers
["K031637","K042053"]
PMA numbers
not on file
Product code
MQB
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register