Skip to the content

FDA Medical Device Recalls (openFDA)

Roche Diagnostics Corporation: cobas p 612 pre-analytical system; Pre-analytical sample handling that includes de-capping, Aliquotting and sorting of samples for analysis.

Recall number
Z-1090-2017
Recalling firm
Roche Diagnostics Corporation
Product
cobas p 612 pre-analytical system; Pre-analytical sample handling that includes de-capping, Aliquotting and sorting of samples for analysis.
Status
Terminated
Initiated
2016-12-27
Posted
2017-01-24
Terminated
2017-08-04
Reason
not on file
Root cause
Software design
Action
On 12/27/2016, URGENT MEDICAL DEVICE CORRECTION notifications were sent to the affected Consignees via courier fax or email. The recall notifications included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. The cobas p 612 pre-analytical system (63x) software is being revised to correct the aspiration process. We anticipate this new version (2.0.4) to be available during Quarter 1 2017. For now, discontinue using disposable non-filter tips for Aliquotting on your cobas p 612 pre-analytical system (63x). You will be contacted by a Roche Field Engineering Specialist (FES) about modifying your system to allow for the use of disposable filter tips. Filter tips will be provided at the time of service. Actions Required " Immediately discontinue using disposable non-filter tips for Aliquotting on your cobas p 612 pre-analytical system (63x). " Complete the enclosed faxback form and fax it to number 1-877-906-8982 or email it to Indianapolis.cd_tsc@roche.com. " Provide copies of this UMDC to other clinicians who may need to be aware of this software issue. " If your facility has distributed the affected product to another site, please ensure this UMDC is provided to that site. " File this UMDC for future reference. Please contact the Roche Support Network Customer Support Center at 1-800-428-2336, 24 hours a day/7 days a week, if you have questions about the information contained in this UMDC.
Quantity
5
Distribution
TX and WA
Lot, serial or model codes
Serial Number 63004316 63004416 63004516 63002816 63002916
510(k) numbers
not on file
PMA numbers
not on file
Product code
JQP
Firm FEI
1823260
Firm city
Indianapolis
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register