Zeus Scientific Inc: Sm ELISA Test System Product number 2Z2831G and Product number 43270CE Individual Enzyme-Linked Immunosorbent Assays for Antibodies to Jo-1, SM, SM/RNP, SSA, SSB and Scl-70. Manufactured by Zeus Scientific, Raritan, NJ 08869-0038. Also manufactured for Wampole Laboratories, Princeton, NJ 08540 Semi-quantitative immunoassays for the detection of IgG antibodies to Jo-1, Sm, Sm/RPN (Ro), SSB(La) and Scl-70 in human sera. This device is for in-vitro diagnostic use.

Recall number
Z-1089-2009
Recalling firm
Zeus Scientific Inc
Product
Sm ELISA Test System Product number 2Z2831G and Product number 43270CE Individual Enzyme-Linked Immunosorbent Assays for Antibodies to Jo-1, SM, SM/RNP, SSA, SSB and Scl-70. Manufactured by Zeus Scientific, Raritan, NJ 08869-0038. Also manufactured for Wampole Laboratories, Princeton, NJ 08540 Semi-quantitative immunoassays for the detection of IgG antibodies to Jo-1, Sm, Sm/RPN (Ro), SSB(La) and Scl-70 in human sera. This device is for in-vitro diagnostic use.
Status
Terminated
Initiated
2008-08-19
Posted
2009-03-31
Terminated
2009-04-30
Reason
The Calibrator Value is printed incorrectly on the Sm ELISA Test System, lot # 08041819. The Calibrator Value (CV) printed on the label is 102; the correct CV should read 266.
Root cause
Employee error
Action
Urgent Recall Notice letters were sent out by Federal Express night mail on August 19, 2008 to all direct account. Wampole (Inverness) will notify their customers in a separate letter. Recall letters explained the reason for the recall and that the product does meet the company's criteria when it is used qualitatively to interpret index values or OD rations as outlined in the kit package insert. Customers are advised in the letter that the Calibrator Value of 102 is incorrect and that it should be 266. Instructions to correctly express the specimen results using specimen and calibrator OD values semi-qualitatively as Unit Values and the Calibrator Value (CV) or 266 are included. Letters also state that results for all patients using the incorrect CV of 102 should be recalculated using the correct CV of 266. All letters included a fax back form indicating that they have received the recall notification, have notified customers to correct the CV value for the Semi-Quantitative Conversion and have notified customers that have received the kit to correct the Semi-Quantitative Conversion of Optical Density to AAU/mL for the affected lot number of the Sm ELISA Test System.
Quantity
33 of lot 08041819, 27 of lot 08041811.
Distribution
Nationwide Distribution
Lot, serial or model codes
Lot number 08041819, Exp. 11/2009;, Lot number 08041811, Exp. 11/2009 (Inverness Medical - Wampole)
510(k) numbers
["K941409"]
PMA numbers
not on file
Product code
LJM
Firm FEI
2242436
Firm city
Somerville
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register