Skip to the content

FDA Medical Device Recalls (openFDA)

Elekta, Inc.: Precise Digital Accelerator Intended to be used for radiation therapy treatments of malignant neoplastic diseases, as determined by a licensed medical practitioner.

Recall number
Z-1088-2015
Recalling firm
Elekta, Inc.
Product
Precise Digital Accelerator Intended to be used for radiation therapy treatments of malignant neoplastic diseases, as determined by a licensed medical practitioner.
Status
Terminated
Initiated
2015-01-07
Posted
2015-02-10
Terminated
2016-04-14
Reason
Elekta has identified a batch of securing bolts that fix the Gantry counter frame weight stack may fail.
Root cause
Other
Action
Elekta sent an Urgent Important Field Safety Notice on January 7, 2015, to all affected customers. Elekta recommended that customers replace the M20 x 50 MM bolts marked 'OF' on the gantry counterbalance frame immediately. The digital linear accelerator must not be used until these 'OF' bolts are replaced. Elekta recommended that customers put the machine in a parked position at 0, and attach the gantry locking straps. A Modification kit will be available as soon as possible which contains the correct bolts. Customers were instructed to complete the Important Field Safety Notice Acknowledgement form and return to their local Elekta Office or Representative. Customers with questions were instructed to contact their local Elekta representative. For questions regarding this recall call 404-993-5886.
Quantity
46
Distribution
Worldwide Distribution - USA including AZ, CA, FL, KS, LA, MN, MO, NY, NC and Internationally to Austria, Bulgaria, Canada, China, Germany, Hungary, Italy, Malaysia, Malta, Netherlands, Norway, Russia, Spain, Sweden, Taiwan and Ukraine.
Lot, serial or model codes
153602, 153538, 153561, 153569, 153532, 153556, 153566, 153563, 153564, 153589, 153530, 153531, 153562, 153584, 153615, 153534, 153534, 153535, 153563, 153542, 153542, 153548, 153548, 153558, 153578, 153595, 153699, 153630, 153728, 153680, 153712. 153717, 153723, 153734, 153702, 153715, 153684, 153710, 153577, 153725, 153726, 153691, 153696, 153713, 153724, 153727
510(k) numbers
["K051932"]
PMA numbers
not on file
Product code
IYE
Firm FEI
1037831
Firm city
Atlanta
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register