FDA Medical Device Recalls (openFDA)
Elekta, Inc.: Precise Digital Accelerator Intended to be used for radiation therapy treatments of malignant neoplastic diseases, as determined by a licensed medical practitioner.
- Recall number
- Z-1088-2015
- Recalling firm
- Elekta, Inc.
- Product
- Precise Digital Accelerator Intended to be used for radiation therapy treatments of malignant neoplastic diseases, as determined by a licensed medical practitioner.
- Status
- Terminated
- Initiated
- 2015-01-07
- Posted
- 2015-02-10
- Terminated
- 2016-04-14
- Reason
- Elekta has identified a batch of securing bolts that fix the Gantry counter frame weight stack may fail.
- Root cause
- Other
- Action
- Elekta sent an Urgent Important Field Safety Notice on January 7, 2015, to all affected customers. Elekta recommended that customers replace the M20 x 50 MM bolts marked 'OF' on the gantry counterbalance frame immediately. The digital linear accelerator must not be used until these 'OF' bolts are replaced. Elekta recommended that customers put the machine in a parked position at 0, and attach the gantry locking straps. A Modification kit will be available as soon as possible which contains the correct bolts. Customers were instructed to complete the Important Field Safety Notice Acknowledgement form and return to their local Elekta Office or Representative. Customers with questions were instructed to contact their local Elekta representative. For questions regarding this recall call 404-993-5886.
- Quantity
- 46
- Distribution
- Worldwide Distribution - USA including AZ, CA, FL, KS, LA, MN, MO, NY, NC and Internationally to Austria, Bulgaria, Canada, China, Germany, Hungary, Italy, Malaysia, Malta, Netherlands, Norway, Russia, Spain, Sweden, Taiwan and Ukraine.
- Lot, serial or model codes
- 153602, 153538, 153561, 153569, 153532, 153556, 153566, 153563, 153564, 153589, 153530, 153531, 153562, 153584, 153615, 153534, 153534, 153535, 153563, 153542, 153542, 153548, 153548, 153558, 153578, 153595, 153699, 153630, 153728, 153680, 153712. 153717, 153723, 153734, 153702, 153715, 153684, 153710, 153577, 153725, 153726, 153691, 153696, 153713, 153724, 153727
- 510(k) numbers
- ["K051932"]
- PMA numbers
- not on file
- Product code
- IYE
- Firm FEI
- 1037831
- Firm city
- Atlanta
- Firm state
- GA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28