FDA Medical Device Recalls (openFDA)
Baxter Healthcare Corp.: Baxter 6060 Multi-Therapy Infusion Pump, product code 2M9832, 606000-40, 606000-40L, 606000-40I; Manufactured by an affiliate of Baxter Healthcare Corp., Deerfield, IL 60013 USA, made in Singapore
- Recall number
- Z-1088-05
- Recalling firm
- Baxter Healthcare Corp.
- Product
- Baxter 6060 Multi-Therapy Infusion Pump, product code 2M9832, 606000-40, 606000-40L, 606000-40I; Manufactured by an affiliate of Baxter Healthcare Corp., Deerfield, IL 60013 USA, made in Singapore
- Status
- Terminated
- Initiated
- 2005-06-13
- Posted
- 2005-08-06
- Terminated
- 2007-06-19
- Reason
- There is an overinfusion condition in the Intermittent mode where the pump may continue to deliver at the higher dose rate instead of switching to the lower keep open rate once the first dose is completed.
- Root cause
- Other
- Action
- Baxter sent the 6/13/05 Urgent Device Correction letter to all Baxter 6060 and 6060E Infusion Pump customers to inform them of a software anomaly which allows the pump to overinfuse in the Intermittent mode of the 6060 pump if the following specific sequence of events occur: * the infusion is running in intermittent mode; * the pump is delivering at the Keep Open (KO) rate; * the pump is then powered off and on; * the clinician initiates new programming of the pump by selecting NO at the 'Resume KO Dose 1?' prompt and selecting YES at the 'Yes to Program' prompt; * the clinician powers the pump off and on again; * and the clinician selects YES at the 'Resume KO Dose 1?' prompt. Until the software upgrade is available, Baxter recommended that the accounts notify healthcare providers of the potential for an overinfusion condition in Intermittent mode if the above sequence is performed during an infusion, and to utilize the Lockout mode after programming the pump. Any questions were directed to Baxter''s Medication Delivery Services at 1-800-843-7867. The letter also informed the accounts of two patient deaths due to medication delivery errors resulting from improper programming of the infusion parameters into the 6060 and 6060E pumps, and reports where the infusion parameters were modified by untrained individuals, which may result in overinfusion or underinfusion conditions. The letter stressed that programming of the pump must be performed only by trained clinicians, or healthcare providers trained and deemed competent by trained clinicians or under the direct supervision of trained clinicians; the Lockout mode feature of the pump should be used to prevent unauthorized changes to the pump programming; and the pump security codes, which disable the Lockout mode, must not be released to anyone other than trained healthcare professionals. A copy of the warning from the Operator''s Manual was included with the letter. Any questions were referred to the Center fo
- Quantity
- 34,000 units
- Distribution
- Nationwide and internationally to Belgium, Germany, Finland, France, the United Kingdom, Ireland, Sweden, Norway, Canada, Japan, New Zealand and Australia.
- Lot, serial or model codes
- all serial numbers
- 510(k) numbers
- ["K002679"]
- PMA numbers
- not on file
- Product code
- FRN
- Firm FEI
- 1416980
- Firm city
- Round Lake
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28