Skip to the content

FDA Medical Device Recalls (openFDA)

PHILIPS MEDICAL SYSTEMS: CombiDiagnost R90 1.0 (709030)

Recall number
Z-1087-2024
Recalling firm
PHILIPS MEDICAL SYSTEMS
Product
CombiDiagnost R90 1.0 (709030)
Status
Open, Classified
Initiated
2022-02-11
Posted
2024-02-14
Terminated
not on file
Reason
While performing a fluoroscopy examination, there is a potential that the Radio Fluoroscopy (RF) viewer will also display a previous patient's radiography images. If the issue occurs, there will be differences in image content, image format and image size.
Root cause
Software design
Action
An Important Product Notice (2021-PD-DXR-034) dated February 15, 2022 was issued via Certified Mail. Philips Field Service Engineers will contact customers to schedule an onsite visit to install update software correcting the issue. Customers are asked to follow the provided instructions to ensure there are no pending images before an exam circulate the notice to all users and to complete and return the response form and circulate the notice to all users of the device for awareness. Contact Philips Customer Care Solutions Center 1-800-722-9377 with questions.
Quantity
86 units
Distribution
US Nationwide distribution in the states of CA, FL, GA, IL, IN, KS, MD, ME, MI, MN, NC, NJ, NY, TX, VA, WA.
Lot, serial or model codes
Model: R90 1.0 (709030); UDI DI: 00884838076747; Software version 1.1.1 or higher; Serial numbers: 10000000, 10000002, 10000003, 10000004, 10000006, 10000011, 10000012, 10000020, 10000024, 10000027, 10000049, 10000050, 10000065, 10000066, 10000067, 10000069, 10000078, 10000081, 10000099, 10000103, 10000107, 10000140, 10000145, 10000152, 10000180, 10001001, 10001010, 10001095, 591362, SN16000009, SN17000003, SN17000007, SN17000015, SN17000022, SN17000111, SN17000114, SN17000118, SN17000120, SN17000121, SN17000122, SN17000123, SN17000128, SN17000129, SN17000130, SN17000145, SN17000146, SN17000147, SN17000148, SN17000153, SN17000157, SN17000163, SN17000165, SN17000166, SN18000003, SN18000004, SN18000010, SN18000021, SN18000025, SN18000033, SN18000037, SN18000038, SN18000044, SN18000045, SN18000053, SN18000054, SN18000057, SN18000060, SN18000068, SN18000069, SN18000079, SN19000004, SN19000006, SN19000011, SN19000020, SN19000034, SN19000036, 27, 40, 41, 42, 77, 101, 131, 155, 180, 190;
510(k) numbers
["K163210"]
PMA numbers
not on file
Product code
JAA
Firm FEI
3017726341
Firm city
Cambridge
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register