Skip to the content

FDA Medical Device Recalls (openFDA)

Baxter Healthcare Corp.: Baxter Sabraset 6060 Administration Set with Pre-attached 100 mL Bag and Cassette, product code 560500-100; Manufactured for Baxter Healthcare Corporation, Deerfield, IL 60015 U.S.A., Made in Mexico

Recall number
Z-1087-05
Recalling firm
Baxter Healthcare Corp.
Product
Baxter Sabraset 6060 Administration Set with Pre-attached 100 mL Bag and Cassette, product code 560500-100; Manufactured for Baxter Healthcare Corporation, Deerfield, IL 60015 U.S.A., Made in Mexico
Status
Terminated
Initiated
2005-05-04
Posted
2005-08-06
Terminated
2006-07-21
Reason
The inlet and outlet lines of the administration set may be reversed, which would result in reverse flow.
Root cause
Other
Action
Baxter sent Urgent Device Correction letters dated 5/4/05 to their customers, to the attention of the Director of Nursing, via first class mail on the same date. The accounts were informed of the one complaint of reverse assembly of the administration set, and were provided with diagrams showing the configuration of a correct assembly and an incorrect assembly of the sets. The accounts were instructed to inspect the sets carefully before use to assure correct assembly. The accounts were requested to destroy any misassembled sets found and complete the certificate of destruction enclosed with the letter, faxing it back to Baxter at 847-270-5457 to receive credit for the destroyed sets.
Quantity
13,229 units
Distribution
Nationwide and internationally to Australia, Belgium, Canada and Japan.
Lot, serial or model codes
product code 560500-100, lots 417519, 417520, 418482, 418483, 418745
510(k) numbers
["K002679"]
PMA numbers
not on file
Product code
FPA
Firm FEI
1416980
Firm city
Round Lake
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register