Stryker Spine: Specialty Straight Static Inserter Assembly Product Usage: This instrument is a part of the Navigator product brand . It is used for implant insertion in to the disc space. The AVS navigator Peek Spacer System offers implants that are interbody fusion devices intended for use as an aid in spinal fixation. These hollow implants are offered in a variety of lengths, heights and lordotic angles designed to adapt to different patient anatomies. They have serrations on the superior and inferior surfaces designed for multidirectional fixation and ergonomically shaped anterior edges and flat posterior edges.

Recall number
Z-1086-2014
Recalling firm
Stryker Spine
Product
Specialty Straight Static Inserter Assembly Product Usage: This instrument is a part of the Navigator product brand . It is used for implant insertion in to the disc space. The AVS navigator Peek Spacer System offers implants that are interbody fusion devices intended for use as an aid in spinal fixation. These hollow implants are offered in a variety of lengths, heights and lordotic angles designed to adapt to different patient anatomies. They have serrations on the superior and inferior surfaces designed for multidirectional fixation and ergonomically shaped anterior edges and flat posterior edges.
Status
Terminated
Initiated
2013-12-18
Posted
2014-02-25
Terminated
2014-08-25
Reason
Stryker has received complaints from customers relating to fracture along the weld that binds the handle to the shaft of the static inserter.
Root cause
Device Design
Action
Stryker sent Urgent Medical Device Removal Notification letter and to their customers via email on 12/18/2013. The letter identified the affected product, problem and actions to be taken. Customers were instructed to examine their inventory, return affected product using enclosed pre-paid mailing label, complete Customer Response Form and fax to Regulatory Compliance at 855-632-9049 or mail it to Spine-RegulatoryActions@Stryker.com. For questions call 201-760-8287, 201-760-8298 or 201-760-8041.
Quantity
21 units
Distribution
US Nationwide Distribution
Lot, serial or model codes
D109140 Catalog # IS3009AVS Lot Serial # 3664, 3759, 3843, 3913, 4079, 4195 and 4242
510(k) numbers
not on file
PMA numbers
not on file
Product code
LXH
Firm FEI
3004024955
Firm city
Allendale
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register