Skip to the content

FDA Medical Device Recalls (openFDA)

Thommen Medical, AG: SPI EASY Abutment. Endosseous dental implant abutment.

Recall number
Z-1086-2007
Recalling firm
Thommen Medical, AG
Product
SPI EASY Abutment. Endosseous dental implant abutment.
Status
Terminated
Initiated
2007-04-19
Posted
2007-07-24
Terminated
2008-05-12
Reason
The undercut on the SPI EASY abutments which engages the stabilization ring of the implant is missing. Although the abutments can still be placed on the implant by the user the parts do not fulfill the quality expectations of the firm.
Root cause
Other
Action
Letter 5/14/2007. Customers are requested to check their current inventory for the recall product. If found they are to return the product for replacement. If they have implanted the product they are to check patient x-rays for the defect. If defects are found they are to do a restoration.
Quantity
84
Distribution
Nationwide.
Lot, serial or model codes
#4_03_700 Batch 8007 REF: 1.03.700 LOT: 5121
510(k) numbers
["K023645"]
PMA numbers
not on file
Product code
DZE
Firm FEI
3003836985
Firm city
Waldenburg
Firm state
not on file
Firm country
Switzerland

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register