FDA Medical Device Recalls (openFDA)
Integra LifeSciences Corporation: The MAYFIELD Ultra 360 Base Unit for head positioner, one unit per package. Product Usage: This product provides the ability to position and hold a patient head while the patient is in the prone, supine, lateral or park-bench and sitting positions. It is intended to be used during diagnostic examinations or surgical procedures where a rigid support between the surgical table and a headrest or skull clamp is necessary and positional freedom is required.
- Recall number
- Z-1084-2015
- Recalling firm
- Integra LifeSciences Corporation
- Product
- The MAYFIELD Ultra 360 Base Unit for head positioner, one unit per package. Product Usage: This product provides the ability to position and hold a patient head while the patient is in the prone, supine, lateral or park-bench and sitting positions. It is intended to be used during diagnostic examinations or surgical procedures where a rigid support between the surgical table and a headrest or skull clamp is necessary and positional freedom is required.
- Status
- Terminated
- Initiated
- 2014-12-19
- Posted
- 2015-02-10
- Terminated
- 2017-02-22
- Reason
- Through an investigation of complaints, Integra found that there is the possibility for the locking handle mechanism of the Base Units to fail during use.
- Root cause
- Nonconforming Material/Component
- Action
- Integra sent an Urgent - Voluntary Medical Device Recall letter dated December 19, 2014 to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to take the following actions: 1. Review your MAYFIELD equipment and determine if you have any MAYFIELD Ultra 360 Base Units 2. Once identified we recommend that you discontinue use and remove them from service until they are repaired by an authorized Integra Repair Center 3. Complete the attached Recall Acknowledgement and Repair Form 4. Upon receipt of the Recall Acknowledgement and Repair Form, Customer Service will contact you and provide an RMA number and instructions for returning the productor questions call 609-936-6822. For questions contact Customer Service: USA 800-654-2873 Outside USA 609-936-5400
- Quantity
- 1053 Units
- Distribution
- Worldwide Distribution - US Nationwide in the states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV, WY, and DC. (25 government consignees) and in the foreign countries: CANADA, ALGERIA, ARAB EMIRATES, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, CHILE, CHINA, COLUMBIA, CYPRESS, DENMARK, DOMINICAN REPUBLIC, FRANCE, GERMANY, GREAT BRITAN, HONG KONG, HUNGARY, INDIA, INDONESIA, ISRAEL, ITALY, JAPAN , JORDAN, KOREA, LIBYA, MACEDONIA, MAYLASIA, MEXICO, NETHERLANDS, POLAND, PORTUGAL, PUERTO RICO, ROMANIA, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SPAIN, SRI LANKA, TAIWAN, THAILAND, and TURKEY.
- Lot, serial or model codes
- All Lot numbers of Mayfield Ultra 360 Base Units, Catalogue #'s A2009 and A2029. Affected Lot Numbers For Catalogue No. A2009 73149, 73355, 73356, 74396, 74730, 74398, 74399, 75796, 75816, 74523, 75838, 75859, 74524, 75873, 74890, 75939, 74892, 76420, 74893, 74894, 74895, 76156, 77225, 76157, 76158, 77831, 76159, 77957, 76160, 78012, 76161, 78548, 77292, 78666, 79067, 79068, 79069, 79071, 80013, 79073, 81782, 79074, 81918, 79075, 79748, 80094, 82194, 80095, 80668, 80669, 84171, 80671, 84288, 82736, 82737, 82738, 85896, 82739, 86400, 82740, 82741, 82742, 84732, 84733, 84734, 84735, 85693, 86817, 86818, 86819, 86820, 86821, 86822, 89453, 89455, 89456, 89457, 92296, 92297, 92298, 92299, 92300, 92301, 94653, 94654, 94655, 99798, 98558, 98559, 98560, 98561, 98562, 94658, 103127, 103128, 103129, 103130, 103131, 104890, 114978, 115809, E19251, 123051 Affected Lot Numbers For Catalogue No. A2029 84180, 84487, 85519, 85815, 85631, 84887, 84888, 86533, 86534, 84889, 86828, 86829, 89302, 86830, 86831, 86832, 86833, 86834, 89534, 89535, 89536, 89537, 89538, and 94167.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- FWZ
- Firm FEI
- 3004608878
- Firm city
- Cincinnati
- Firm state
- OH
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28