Skip to the content

FDA Medical Device Recalls (openFDA)

Ormco/Sybronendo: Bracket Buccal Tube U7L DB/UL22 Roth MT HK 0/-14/0 Product Usage: Bracket Buccal Tubes are intended to hold in orthodontic wires used to move teeth during orthodontic treatment

Recall number
Z-1084-2013
Recalling firm
Ormco/Sybronendo
Product
Bracket Buccal Tube U7L DB/UL22 Roth MT HK 0/-14/0 Product Usage: Bracket Buccal Tubes are intended to hold in orthodontic wires used to move teeth during orthodontic treatment
Status
Terminated
Initiated
2013-03-08
Posted
2013-04-10
Terminated
2013-09-04
Reason
Ormco Corporation is voluntarily recalling one (1) lot of Bracket Buccal Tubes as a result of a packaging error. More specifically, the Bracket Buccal Tubes product package correctly identifies the product as gingival (0) offset brackets, but actually contains distal (10) offset brackets. Use of the affected product may cause unintended rotation of the molars. Although reversible, a delay in
Root cause
Packaging
Action
Ormco sent an Urgent Medical Device Recall letter dated March 8, 2013 to all affected customers, via USPS 1st class mail. The letter identified the affected product, problem and actions to be taken. The consignees were instructed to complete the Return Form and return any affected product identified in stock. For questions contact Ormco Customer Care.
Quantity
8210
Distribution
Worldwide Distribution - USA (nationwide) including the country of EU.
Lot, serial or model codes
Part Number  438-0184   Lot Number 051063597
510(k) numbers
not on file
PMA numbers
not on file
Product code
DZD
Firm FEI
2016150
Firm city
Glendora
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register