FDA Medical Device Recalls (openFDA)
Centurion Medical Products Corporation: CIRCUMCISION TRAY
- Recall number
- Z-1082-2018
- Recalling firm
- Centurion Medical Products Corporation
- Product
- CIRCUMCISION TRAY
- Status
- Terminated
- Initiated
- 2017-10-23
- Posted
- not on file
- Terminated
- 2018-06-21
- Reason
- Convenience Kit contains a PVP Prep Pad, received by a supplier, that does not meet the iodine assay level required to support the 36 month shelf life.
- Root cause
- Nonconforming Material/Component
- Action
- On October 16, 2017 a PRODUCT CORRECTION letter was issued to customers notifying them that if they had any kits in their inventory, Centurion will supply labels to affix to the front of each kit instructing the user to discard the prep pad upon opening the kit. Other components of the kit may continue to be used and are NOT within the scope of this recall.
- Quantity
- not on file
- Distribution
- Nationally
- Lot, serial or model codes
- CIT6020 CIT6120 Lot codes: 2015081790 2015112090 2016011190 2016022390 2016031690 2016042690 2016061590 2016080890 2016090690 2016100790 2017010990 2017020990 2017040490 2017051590 2017061690 2015111390 2016012190 2016022390 2016033190 2016060690 2016071890 2016092290 2016111590 2017020990 2017030790 2017041390 2017051590
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- LRP
- Firm FEI
- 3008403546
- Firm city
- Williamston
- Firm state
- MI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28