Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer Biomet, Inc.: Biomet Small Diameter Cement Plug Inserter-Sterile, single-use accessory instrumentation to be used during surgical procedures utilizing the Small Diameter Cement Plug implant devices Item Number: 414989

Recall number
Z-1081-2019
Recalling firm
Zimmer Biomet, Inc.
Product
Biomet Small Diameter Cement Plug Inserter-Sterile, single-use accessory instrumentation to be used during surgical procedures utilizing the Small Diameter Cement Plug implant devices Item Number: 414989
Status
Terminated
Initiated
2019-02-28
Posted
not on file
Terminated
2020-04-23
Reason
Lack of an adequate sterilization validation
Root cause
Process design
Action
Zimmer Biomet issued recall letter on 2/28/19 via FedEx and email identifying reason for recall to the consumer/user level, including any intermediate wholesale or retail consignees. Distributors responsibilities include locating and removing the product in their territory, as well as identifying hospitals and surgeons who have previously used the product. " Distributors will return on-hand product to Zimmer Biomet and ensure all of their products are accounted for using the form provided in the letter. " Surgeons provided with a letter that identifies the issue and their responsibilities, which include: - Following any specified patient monitoring instructions - Returning Certificate of Acknowledgment to Zimmer Biomet. " Hospital risk managers provided with a letter identifying the issue and their responsibilities. Questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday
Quantity
89 units
Distribution
Nationwide
Lot, serial or model codes
All lots
510(k) numbers
["K033280"]
PMA numbers
not on file
Product code
KIH
Firm FEI
1825034
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register