FDA Medical Device Recalls (openFDA)

Integra LifeSciences Corp: Suturable DuraGen Dural Regeneration Matrix; Non-Pyrogenic; Sterility guaranteed unless package is damaged or opened. Rx only. Manufacturer: Integra LifeSciences Corporation, Plainsboro, NJ 08536 Suturable DuraGen is an absorbable implant for repair of dural defects. It is indicated as a dural substitute for the repair of dura matter.

Recall number
Z-1081-2010
Recalling firm
Integra LifeSciences Corp
Product
Suturable DuraGen Dural Regeneration Matrix; Non-Pyrogenic; Sterility guaranteed unless package is damaged or opened. Rx only. Manufacturer: Integra LifeSciences Corporation, Plainsboro, NJ 08536 Suturable DuraGen is an absorbable implant for repair of dural defects. It is indicated as a dural substitute for the repair of dura matter.
Status
Terminated
Initiated
2010-01-13
Posted
2010-03-11
Terminated
2010-03-18
Reason
One lot of Suturable DuraGen Dural Regeneration Matrix (# 1091575) did not pass stability testing at certain time points for Suture Retention Strength and Shrink Temperature specifications.
Root cause
Other
Action
An "URGENT PRODUCT RECALL NOTIFICATION" dated January 13, 2010, was sent via Federal Express to customers. The notification described the product, problem, hazard involved and action to be taken by customers. In the event that the customers have in their possession any products from the specific lot number 1091576, the customers are being ask to please immediately quarantine all stock on hand and contact their local Integra Sales representative for further assistance and for arrangements of product return. The customers are being ask to return the enclosed "Recall Acknowledge and Return Form" indicating whether or not they possess any of the affected product identified. Integra will replace any product affected by this Recall that is returned. Please feel free to contact me at 609-936-5407 should you have any additional questions.
Quantity
94
Distribution
Nationwide distribution:Ca, NY, TX, CO, VA, SC, IA, MO, IL, LA, OR, OH, WI, FL, DC, DE, MD, MN.
Lot, serial or model codes
Catalog Number DURS3391; Lot number 1091575
510(k) numbers
["K043427"]
PMA numbers
not on file
Product code
GXQ
Firm FEI
1121308
Firm city
Plainsboro
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register